WAVESCAN WS1 0070-0150

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2015-06-18 for WAVESCAN WS1 0070-0150 manufactured by Abbott Medical Optics.

Event Text Entries

[5868345] Account reported that patient had a decentered ablation with interrupted procedure in right eye and that as a result the patient experienced loss of best corrected visual acuity equal to 2 lines in his right eye. Patient history: pre-op manifest refraction on (b)(6) 2015: right eye +4. 50 -0. 50 x 040 bcva 20/20, left eye +3. 25 -0. 75 x 146 bcva 20/15. Patient ilasik treatment was (b)(6) 2015. Desired correction: right eye: +4. 72 -0. 64 x 034 with -0. 20 physician adjustment, left eye: +3. 46 -0. 63 x 146 with -0. 20 physician adjustment. Post op visit for (b)(6) 2015 bcva: right eye 20/30. Last post office visit was on (b)(6) 2015 and uncorrected visual acuity was: right eye 20/30, left eye 20/25, with no refraction in both eyes. Surgeon commented that he has observed the wavescan refraction has been 0. 50 diopter less than manifest refraction. Surgeon requested to have field service check out both systems, the wavescan and star excimer s4ir. This is 2 of 2 reports.
Patient Sequence No: 1, Text Type: D, B5


[13494710] Application support manager (asm) discussed with surgeon that wavescan refraction sphere will be 0. 50 diopter less when compared to manifest refraction sphere due to monochromatic aberrations unless patient is accommodating. Plastics dated (b)(6) 2015 were checked and found with no artifacts. Customer did not have plastic for (b)(6) 2014. Wavescan daily calibration performed by the asm was found within specifications and a system checked by field service engineer (fse) found system within specifications. All pertinent information available to abbott medical optics has been submitted. Placeholder.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3006695864-2015-00301
MDR Report Key4854830
Report Source05,06
Date Received2015-06-18
Date of Report2015-05-22
Date of Event2015-05-22
Date Mfgr Received2015-05-22
Date Added to Maude2015-06-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMRS. VALERIE SEDZICKI
Manufacturer Street1700 EAST ST. ANDREW PLACE
Manufacturer CitySANTA ANA CA 92705
Manufacturer CountryUS
Manufacturer Postal92705
Manufacturer Phone7142478567
Manufacturer G1ABBOTT MEDICAL OPTICS INC.
Manufacturer Street510 COTTONWOOD DRIVE
Manufacturer CityMILPITAS CA 95035
Manufacturer CountryUS
Manufacturer Postal Code95035
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWAVESCAN WS1
Generic NameREFRACTIVE MEASUREMENT
Product CodeHKO
Date Received2015-06-18
Model Number0070-0150
OperatorPHYSICIAN
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MEDICAL OPTICS
Manufacturer AddressSANTA ANA CA US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-06-18

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.