VERSACELL SAMPLE MANAGEMENT SYSTEM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2015-06-19 for VERSACELL SAMPLE MANAGEMENT SYSTEM manufactured by Siemens Healthcare Diagnostics Inc..

Event Text Entries

[5806852] The operator of a versacell sample management system incorrectly assigned a sample identification number to a patient sample. Patient sample (b)(6) was initially loaded into the sample drawer, scanned, queried, and delivered to a dimension instrument for testing. The operator then front loaded a different patient sample and assigned the same patient sample identification (b)(6) to the specimen, which was delivered to an alternate dimension instrument for testing. The incorrect results of the specimen which was front loaded were reported to the physician(s). The results were corrected and reported to the physician(s). There are no reports of patient intervention or adverse health consequences due to the incorrect assignment of a patient identification number.
Patient Sequence No: 1, Text Type: D, B5


[13378025] The customer contacted the siemens customer care center (ccc). Ccc reviewed the versacell log and confirmed that the operator front loaded the sample with an incorrect identification number. The cause of the incorrect assignment of a patient identification number was due to user error. The instrument is performing within specifications. No further evaluation of the device is required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2247117-2015-00037
MDR Report Key4856119
Report Source05,06
Date Received2015-06-19
Date of Report2015-05-27
Date of Event2015-05-27
Date Mfgr Received2015-05-27
Date Added to Maude2015-07-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMARGARITA KARAN
Manufacturer Street511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145243105
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC
Manufacturer Street62 FLANDERS BARTLEY RD
Manufacturer CityFLANDERS NJ 07836
Manufacturer CountryUS
Manufacturer Postal Code07836
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVERSACELL SAMPLE MANAGEMENT SYSTEM
Generic NameCLINICAL CHEMISTRY GENERAL PURPOSE LABORATORY EQUIPMENT
Product CodeCDN
Date Received2015-06-19
Model NumberVERSACELL SAMPLE MANAGEMENT SYSTEM
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Address62 FLANDERS BARTLEY RD FLANDERS NJ 07836 US 07836


Patients

Patient NumberTreatmentOutcomeDate
10 2015-06-19

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.