AMS ACTICON NEOSPHINCTER

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2015-06-19 for AMS ACTICON NEOSPHINCTER manufactured by American Medical Systems (mn).

Event Text Entries

[5871878] It was reported that the patient is scheduled to have his artificial bowel sphincter removed and replaced on (b)(6) 2015 due to "the pump is no longer working or there is fluid loss. " no patient complications have been reported in relation to this event.
Patient Sequence No: 1, Text Type: D, B5


[13476870] (in order of cuff, pump, balloon): serial #: (b)(4). Catalog #: 72401984, 72402287, 72402106. Expiration date: 4/29/2014, 8/15/2014, 4/02/2015. (in order of cuff, pump, balloon): manufacture date: 5/01/2009, 8/20/2013, 4/15/2013.
Patient Sequence No: 1, Text Type: N, H10


[22569687] Analysis results: the aus 800 device was visually inspected. There was a leak in the cuff pillow at the fillet junction that was the result of fatigue. There was a leak in the pump-cuff tube that was the result of a sharp instrument which resembled a needle hole. The pump and balloon were not functionally tested due to the cuff leak.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2183959-2015-00262
MDR Report Key4857333
Report Source05
Date Received2015-06-19
Date of Report2015-06-12
Date Mfgr Received2015-07-09
Date Added to Maude2015-06-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMRS. SHARON ZURN
Manufacturer Street10700 BREN ROAD W
Manufacturer CityMINNETONKA MN 55343
Manufacturer CountryUS
Manufacturer Postal55343
Manufacturer Phone9529306000
Manufacturer G1AMERICAN MEDICAL SYSTEMS (MN)
Manufacturer Street10700 BREN ROAD W
Manufacturer CityMINNETONKA MN 55343
Manufacturer CountryUS
Manufacturer Postal Code55343
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAMS ACTICON NEOSPHINCTER
Generic NameIMPLANTED FECAL INCONTINENCE DEVICE
Product CodeMIP
Date Received2015-06-19
Returned To Mfg2015-07-09
OperatorLAY USER/PATIENT
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerAMERICAN MEDICAL SYSTEMS (MN)
Manufacturer Address10700 BREN ROAD W MINNETONKA MN 55343 US 55343


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Required No Informationntervention 2015-06-19

© 2026 FDA.report