MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2015-06-23 for URISYS 1100 03617556001 manufactured by Roche Diagnostics.
[18660746]
Reporter stated that a patient urine sample was tested on the urisys 1100 which reported a normal result for glucose and negative results for protein and erythrocytes. A second test, using the same sample on the urisys 1100, reported results of 250 mg/dl for glucose, 500 mg/dl for protein, and 50 erythrocytes/microliter. Reporter did not know which value was correct. No modification to patient therapy was reported based on values. No adverse event was reported. The manufacturer requested the return of the suspect product for evaluation.
Patient Sequence No: 1, Text Type: D, B5
[18952206]
It is not known if the initial reporter has or intends to report the event to fda.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 1823260-2015-03697 |
MDR Report Key | 4861896 |
Report Source | 06 |
Date Received | 2015-06-23 |
Date of Report | 2015-07-17 |
Date of Event | 2015-05-20 |
Date Mfgr Received | 2015-05-28 |
Date Added to Maude | 2015-06-23 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 3 |
Manufacturer Contact | NA MICHAEL LESLIE |
Manufacturer Street | 9115 HAGUE ROAD NA |
Manufacturer City | INDIANAPOLIS IN 46250 |
Manufacturer Country | US |
Manufacturer Postal | 46250 |
Manufacturer Phone | 3175214343 |
Manufacturer G1 | ROCHE DIAGNOSTICS GMBH |
Manufacturer Street | SANDHOFERSTRASSE 116 NA |
Manufacturer City | MANNHEIM (BADEN-WURTTEMBERG) 68305 |
Manufacturer Country | GM |
Manufacturer Postal Code | 68305 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 3 |
Brand Name | URISYS 1100 |
Generic Name | AUTOMATED URINE ANALYZER |
Product Code | KHE |
Date Received | 2015-06-23 |
Returned To Mfg | 2015-07-03 |
Model Number | NA |
Catalog Number | 03617556001 |
Lot Number | NA |
ID Number | NA |
Device Availability | R |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2015-06-23 |