MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05 report with the FDA on 2003-09-30 for STAT PROFILE M MGH SENSOR I MAGNESIUM ELECTRODE 20394 manufactured by Nova Biomedical.
        [295167]
General mgr reports that hosp using a stat profile m analyzer found recovery of mg++ results of pts lower than a nova 8. The average level by nova 8 was 0. 52 mmol/2 vs 0. 35 mmol/l on stat profile m.
 Patient Sequence No: 1, Text Type: D, B5
| Report Number | 1219029-2003-00002 | 
| MDR Report Key | 486714 | 
| Report Source | 01,05 | 
| Date Received | 2003-09-30 | 
| Date of Report | 2003-09-29 | 
| Date of Event | 2003-05-06 | 
| Date Mfgr Received | 2003-05-06 | 
| Device Manufacturer Date | 2002-07-01 | 
| Date Added to Maude | 2003-10-06 | 
| Event Key | 0 | 
| Report Source Code | Manufacturer report | 
| Manufacturer Link | Y | 
| Number of Patients in Event | 0 | 
| Adverse Event Flag | 3 | 
| Product Problem Flag | 3 | 
| Reprocessed and Reused Flag | 0 | 
| Health Professional | 3 | 
| Initial Report to FDA | 3 | 
| Report to FDA | 3 | 
| Event Location | 0 | 
| Manufacturer Contact | PAUL MAC DONALD | 
| Manufacturer Street | 200 PROSPECT ST | 
| Manufacturer City | WALTHAM MA 02454 | 
| Manufacturer Country | US | 
| Manufacturer Postal | 02454 | 
| Manufacturer Phone | 7818940800 | 
| Manufacturer G1 | * | 
| Manufacturer Street | * | 
| Manufacturer City | * | 
| Manufacturer Country | * | 
| Single Use | 3 | 
| Remedial Action | RC | 
| Previous Use Code | 3 | 
| Removal Correction Number | NA | 
| Event Type | 3 | 
| Type of Report | 3 | 
| Brand Name | STAT PROFILE M MGH SENSOR | 
| Generic Name | IONIZED MAGNESIUM SENSOR | 
| Product Code | CFA | 
| Date Received | 2003-09-30 | 
| Model Number | I MAGNESIUM ELECTRODE | 
| Catalog Number | 20394 | 
| Lot Number | 207152 | 
| ID Number | * | 
| Operator | HEALTH PROFESSIONAL | 
| Device Availability | Y | 
| Device Eval'ed by Mfgr | Y | 
| Implant Flag | N | 
| Date Removed | A | 
| Device Sequence No | 1 | 
| Device Event Key | 475600 | 
| Manufacturer | NOVA BIOMEDICAL | 
| Manufacturer Address | 200 PROSPECT ST. WALTHAM MA 02454 US | 
| Baseline Brand Name | STAT PROFILE M MGH SENSOR | 
| Baseline Generic Name | IONIZED MAGNESIUM SENSOR | 
| Baseline Model No | I MAGNESIUM ELE | 
| Baseline Catalog No | 20394 | 
| Baseline ID | * | 
| Patient Number | Treatment | Outcome | Date | 
|---|---|---|---|
| 1 | 0 | 2003-09-30 |