MOBIUS POWER *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-03-23 for MOBIUS POWER * manufactured by Enovate Medical.

Event Text Entries

[21118697] A nurse (rn) reported to engineering a burning electrical smell permeating throughout the wing. Engineering and security responded, searched and located a workstation on wheels (wow) with a flickering charger display and the strong smell of smoke next to an unoccupied patient bed. The unit was unplugged and transported to the ground floor utility area for storage until repaired. Several minutes later the utility area smoke alarm activated and the team responded to find the lithium-ion battery emitting smoke. The wow was transported to the back loading dock area and held to determine source of smoke and evaluation. The wow's batteries then began to spark and emit large flames several times. The team then sprayed the wow with a dry soda based extinguisher and placed the device into an empty metallic bin where it continued to flame up and melt the battery and workstation. City fire was dispatched and biomedical was contacted to determine next steps approximately an hour and a half later. City fire contained equipment until there was no further burning of the device and handed it off to hospital biomedical and it collaborated to contact manufacturer and submit adverse event report. No patient harm or care area evacuation occurred due to this event. The manufacturer is determining further plan of action on retrieving devices for analysis and report.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number4881673
MDR Report Key4881673
Date Received2015-03-23
Date of Report2015-03-23
Date of Event2015-03-23
Report Date2015-03-23
Date Reported to FDA2015-03-23
Date Reported to Mfgr2015-07-01
Date Added to Maude2015-07-01
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameMOBIUS POWER
Generic NameWORKSTATION, MOBILE, BATTERY OPERATED
Product CodeFCO
Date Received2015-03-23
Model NumberMOBIUS POWER
Catalog Number*
Lot Number*
ID Number*
Device AvailabilityY
Device Age4 MO
Device Sequence No1
Device Event Key0
ManufacturerENOVATE MEDICAL
Manufacturer Address1152 PARK AVENUE MURFREEBORO TN 37129 US 37129

Device Sequence Number: 2

Brand NameMOBIUS POWER
Generic NameBATTERY, LITHIUM ION
Product CodeFCO
Date Received2015-03-23
Model NumberMOBIUS POWER 26 AH
Catalog NumberHC20-MB-1B06
Lot Number*
ID Number*
Device AvailabilityY
Device Age4 MO
Device Sequence No2
Device Event Key0
ManufacturerENOVATE MEDICAL
Manufacturer Address1152 PARK AVENUE MURFREEBORO TN 37129 US 37129


Patients

Patient NumberTreatmentOutcomeDate
10 2015-03-23

© 2026 FDA.report