HUDSON BREATHING CIRCUIT,HEATED WIRE NIV 780-98

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06,07 report with the FDA on 2015-07-01 for HUDSON BREATHING CIRCUIT,HEATED WIRE NIV 780-98 manufactured by Teleflex Medical.

Event Text Entries

[6043775] The customer alleges that there is a slight tear found in the circuit tubing. No patient injury or harm reported.
Patient Sequence No: 1, Text Type: D, B5


[13307391] (b)(4). The device sample was not returned for evaluation at the time of this report.
Patient Sequence No: 1, Text Type: N, H10


[23703634] (b)(4). The reported complaint that there was a slight tear found in the circuit tube during use was confirmed through visual inspection of the returned sample. The tear on the tube was identified and noted on the expiratory limb. A dhr review could not be conducted because the lot number was not provided. The investigation found no evidence to suggest a manufacturing related cause, therefore, the root cause is undetermined.
Patient Sequence No: 1, Text Type: N, H10


[23703635] The customer alleges that there is a slight tear found in the circuit tubing. No patient injury or harm reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004365956-2015-00185
MDR Report Key4881972
Report Source05,06,07
Date Received2015-07-01
Date of Report2015-06-16
Date of Event2015-06-15
Date Mfgr Received2015-07-24
Date Added to Maude2015-07-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRESPIRATORY THERAPIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactKATHARINE TARPLEY
Manufacturer Street3015 CARRINGTON MILL BLVD
Manufacturer CityMORRISVILLE NC 27560
Manufacturer CountryUS
Manufacturer Postal27560
Manufacturer Phone9194334854
Manufacturer G1TELEFLEX MEDICAL
Manufacturer StreetPARQUE INDUSTRIAL FINSA
Manufacturer CityNUEVO LAREDO 88275
Manufacturer CountryMX
Manufacturer Postal Code88275
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHUDSON BREATHING CIRCUIT,HEATED WIRE NIV
Generic NameBREATHING CIRCUIT
Product CodeCAG
Date Received2015-07-01
Returned To Mfg2015-07-06
Catalog Number780-98
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTELEFLEX MEDICAL
Manufacturer AddressRESEARCH TRIANGLE PARK NC


Patients

Patient NumberTreatmentOutcomeDate
10 2015-07-01