POSSIS 4320

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 08 report with the FDA on 2015-07-01 for POSSIS 4320 manufactured by External Manufacturer.

Event Text Entries

[16560580] Boston scientific received information that the patient received 28 inappropriate shocks due to noise and oversensing of the right ventricular (rv) lead and resulted in device therapy exhaustion. Boston scientific technical services (ts) was contacted and advised that the integrity of the lead system is in question. The leads were capped and replaced. No adverse patient effects were reported.
Patient Sequence No: 1, Text Type: D, B5


[16610548] As no further information concerning this report is expected, our investigation is complete. This investigation will be updated should further information be provided.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2124215-2015-06455
MDR Report Key4883139
Report Source08
Date Received2015-07-01
Date of Report2015-04-29
Date of Event2015-04-24
Date Mfgr Received2015-04-29
Date Added to Maude2015-07-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactSONALI VASEKAR
Manufacturer Street4100 HAMLINE AVE. N
Manufacturer CityST. PAUL MN 55112
Manufacturer Postal55112
Manufacturer Phone6515824786
Manufacturer G1EXTERNAL MANUFACTURER
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePOSSIS
Generic NameIMPLANTABLE LEAD
Product CodeNHW
Date Received2015-07-01
Model Number4320
ID Number---
Device Expiration Date1994-08-01
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerEXTERNAL MANUFACTURER


Patients

Patient NumberTreatmentOutcomeDate
1411. Hospitalization; 2. Life Threatening; 3. Required No Informationntervention 2015-07-01

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.