RUSCH MRI COND FIBER OPTIC LARYN KIT 005852300

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2015-07-01 for RUSCH MRI COND FIBER OPTIC LARYN KIT 005852300 manufactured by Teleflex Medical.

Event Text Entries

[19818290] The customer alleges that the light bundle will not light on the mac 2 blade. Intubation was accomplished with a standard laryngoscope set. No patient injury reported.
Patient Sequence No: 1, Text Type: D, B5


[19867888] (b)(4). The device sample was not returned for evaluation at the time of this report.
Patient Sequence No: 1, Text Type: N, H10


[23950212] (b)(4). The mri kit was returned for evaluation. A visual exam was performed and it was observed that the mac 2 blade was missing. The complaint was reported as the mac 2 blade would not energize when engaged on the handle. The mac 2 blade was not present in the kit; however, during functional testing of all of the returned blades, it was found that two of the blades would not energize: the mac 3, and the mil 1. The small incandescent bulbs which are seated in the focs light bundle were either defective or burned out. Based on the investigation performed, the reported complaint was confirmed. It should be noted that the blade that was reported defective was not returned, but there were defective blades in the kit. The root cause as to why the bulbs burned out could not be determined. A conclusion code could not be found as the complaint was confirmed, however, a root cause was not established.
Patient Sequence No: 1, Text Type: N, H10


[23950213] The customer alleges that the light bundle will not light on the mac 2 blade. Intubation was accomplished with a standard laryngoscope set. No patient injury reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1044475-2015-00215
MDR Report Key4883781
Report Source06
Date Received2015-07-01
Date of Report2015-06-17
Date of Event2014-06-18
Date Mfgr Received2015-07-10
Date Added to Maude2015-07-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactKATHARINE TARPLEY
Manufacturer Street3015 CARRINGTON MILL BLVD
Manufacturer CityMORRISVILLE NC 27560
Manufacturer CountryUS
Manufacturer Postal27560
Manufacturer Phone9194334854
Manufacturer G1TELEFLEX MEDICAL
Manufacturer Street2917 WECK DRIVE
Manufacturer CityRESEARCH TRIANGLE PARK NC 27709
Manufacturer CountryUS
Manufacturer Postal Code27709
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRUSCH MRI COND FIBER OPTIC LARYN KIT
Generic NameMRI LARYNGOSCOPE KIT
Product CodeOGH
Date Received2015-07-01
Returned To Mfg2015-07-01
Catalog Number005852300
Lot Number132601
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTELEFLEX MEDICAL
Manufacturer AddressRESEARCH TRIANGLE PARK NC


Patients

Patient NumberTreatmentOutcomeDate
10 2015-07-01

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.