CUSA EXCEL CUSAEXCEL NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2003-09-23 for CUSA EXCEL CUSAEXCEL NA manufactured by Valleylab Div Tyco Healthcare Gr.

Event Text Entries

[401918] Add'l info rec'd from mfr 11/11/03: the medwatch sent by the customer identified the console, rather than the cusa tip. The cusa tip was not returned to valleylab for eval; however, the handpiece had been returned to valleylab's service dept on july 18th, 2003. The test results found the handpiece to operate properly and within all specs when tested with a spare test tip. The handpiece was disassembled and no conditions were found that would have caused or contributed to the tip breaking. Although the cause of the customer's report cannot be determined, tip breakage is known to occur when the tip is placed in close proximity to electrosurgical energy during use. The tip will produce a bluish color, but again, the tip was not returned for exam. Product labeling cautions the user not to use the tip in close proximity to electrosurgery.
Patient Sequence No: 1, Text Type: D, B5


[21583000] Tip of the cusa (ultrasounic surgical aspirator) handpiece broke while in use. Tip retrieved, no injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW1029632
MDR Report Key488779
Date Received2003-09-23
Date of Report2003-09-16
Date of Event2003-06-21
Date Added to Maude2003-10-16
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA0
Report to FDA0
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameCUSA EXCEL
Generic NameULTRASONIC SURGICAL ASPIRATOR
Product CodeLBK
Date Received2003-09-23
Returned To Mfg2003-06-25
Model NumberCUSAEXCEL
Catalog NumberNA
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key477606
ManufacturerVALLEYLAB DIV TYCO HEALTHCARE GR
Manufacturer Address5920 LONGBOW DR BOULDER CO 803013299 US
Baseline Brand NameCUSA EXCEL CONSOLE
Baseline Generic NameULTRASONIC ASPIRATOR
Baseline Model NoCUSAEXCEL
Baseline Catalog NoCUSAEXCEL
Baseline Shelf Life Contained*
Baseline Shelf Life [Months]*
Baseline PMA FlagN
Baseline 510K PMNY
Premarket NotificationK981262
Baseline PreamendmentN
Baseline TransitionalN
510k ExemptN


Patients

Patient NumberTreatmentOutcomeDate
10 2003-09-23

© 2026 FDA.report