ADVIA CENTAUR CP

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2015-07-07 for ADVIA CENTAUR CP manufactured by Siemens Healthcare Diagnostics Inc..

Event Text Entries

[6018866] Discordant results were obtained on patient samples tested for multiple methods on an advia centaur cp instrument. Discordant results were not reported to the physician(s), as the customer stated that results were not believable. Quality controls were out of range for all methods. It is unknown if repeat testing was performed. There are no reports of patient intervention or adverse health consequences due to the discordant results.
Patient Sequence No: 1, Text Type: D, B5


[14226790] A siemens customer service engineer (cse) was dispatched to the customer site. After evaluation of the instrument and instrument data, the cse replaced the wash 1 solution due to contamination and cleaned the luminometer due to base residue. The cause of the discordant results is unknown. The instrument is performing according to specifications. No further evaluation of the device is required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2432235-2015-00318
MDR Report Key4894270
Report Source05,06
Date Received2015-07-07
Date of Report2015-06-11
Date of Event2015-06-11
Date Mfgr Received2015-06-11
Date Added to Maude2015-08-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMARGARITA KARAN
Manufacturer Street511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145243105
Manufacturer G1STRATEC BIOMEDICAL SWITZERLAND AG
Manufacturer StreetREGISTRATION #: 3008494306 NEUWIESENSTRASSE 4
Manufacturer CityBERINGEN, 8222
Manufacturer CountrySZ
Manufacturer Postal Code8222
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameADVIA CENTAUR CP
Generic NameIMMUNOASSAY ANALYZER
Product CodeMOI
Date Received2015-07-07
Model NumberADVIA CENTAUR CP
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Address511 BENEDICT AVENUE TARRYTOWN NY 10591 US 10591


Patients

Patient NumberTreatmentOutcomeDate
10 2015-07-07

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.