[6542212] It was reported that a dental professional was positioning a pelton and crane lftn dental light for use when the light assembly fell down from the track assembly and hit the patient on the shoulder. There were no injuries reported.
Patient Sequence No: 1, Text Type: D, B5


[13767724] Upon evaluation by the local pelton and crane distributor, it was determined the roll pins were not installed by the distributor during installation/servicing. The roll pins will prevent the light from unscrewing from the pole after installation. The dental light was manufactured over 38 years ago and is past the expected life of the device.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1017522-2015-00013
MDR Report Key4908525
Report Source08
Date Received2015-07-09
Date of Report2015-06-12
Date of Event2015-06-10
Date Mfgr Received2015-06-12
Device Manufacturer Date1977-05-01
Date Added to Maude2015-07-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactFRANK RAY
Manufacturer Street11727 FRUEHAUF DR.
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal28273
Manufacturer Phone7045877227
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLFTN TRACK MOUNTED LIGHT
Generic NameDENTAL LIGHT
Product CodeEAZ
Date Received2015-07-09
Model NumberHA
Catalog NumberLFTN
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerPELTON & CRANE
Manufacturer Address11727 FRUEHAUF DR. CHARLOTTE NC 28273 US 28273


Patients

Patient NumberTreatmentOutcomeDate
10 2015-07-09