THROMBOMAX HS WITH CALCIUM T9686

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 03,08 report with the FDA on 2003-10-20 for THROMBOMAX HS WITH CALCIUM T9686 manufactured by Sigma Aldrich, Co..

Event Text Entries

[331326] Incorrect isi mechnical and optical values were printed on the package insert. Comparison of values: incorrect isi value, mechanical: 1. 01. Optical: 0. 96, correct isi value: mechanical: 1. 25, optical: 1. 17. Isi values are used to adjust prothrombin time measures, which are clinically relevant to clotting time of blood. These clotting times are used with other clinical indicators to assess and/or adjust blood therapeutics, including thinners such as coumadin. The incorrect values of isi may lead to clotting time(s) adjustment lower (less time) than actual, thereby potentially indicating increase dosage of blood thinners. Regarding the malfunction, no such events have been reported to date for this.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1938173-2003-00001
MDR Report Key491234
Report Source03,08
Date Received2003-10-20
Date of Report2003-09-18
Date Mfgr Received2003-09-18
Device Manufacturer Date2002-10-01
Date Added to Maude2003-10-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJESSICA LOVE
Manufacturer Street545 S. EWING AVE
Manufacturer CitySAINT LOUIS MO 63103
Manufacturer CountryUS
Manufacturer Postal63103
Manufacturer Phone8003253010
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Remedial ActionNO
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameTHROMBOMAX HS WITH CALCIUM
Generic NameTHROMBOPLASTIN REAGENT
Product CodeGGO
Date Received2003-10-20
Model NumberNA
Catalog NumberT9686
Lot Number102K6071
ID NumberNA
Device Expiration Date2004-10-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeNA
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key479947
ManufacturerSIGMA ALDRICH, CO.
Manufacturer Address* ST. LOUIS MO * US
Baseline Brand NameTHROMBOMAX HS WITH CALCIUM
Baseline Generic NameTHROMBOPLASTIN REAGENT
Baseline Model NoNA
Baseline Catalog NoT9686
Baseline IDNA


Patients

Patient NumberTreatmentOutcomeDate
10 2003-10-20

© 2024 FDA.report