MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2015-07-10 for ALLSET GOLD A HIGH RES SSP 54010D manufactured by Life Technologies.
        [24646741]
The investigation replicated the customer's observation (b)(6) 2015. The investigation concluded on 06/16/2015 and was unable to determine whether the result is related to the specific dna sample or reflective or incorrect labeling on the reactivity for the allele.
 Patient Sequence No: 1, Text Type: N, H10
        [24646742]
This report is related to customer complaint ci23108 made against allset gold a high res ssp (sku 54010d, lot 012 1608823), since this lot (listed in c. 6. ) contains the same affected primer mix. The customer reported that the reactivity for lane 86 was negative for the a 03:20 allele when the labeling indicates the reactivity should be positive. The labeling discrepancy may lead to potentially mistyping an a 03:20 allele as an a 03:01 allele.
 Patient Sequence No: 1, Text Type: D, B5
| Report Number | 2244574-2015-00092 | 
| MDR Report Key | 4912900 | 
| Report Source | 05 | 
| Date Received | 2015-07-10 | 
| Date of Report | 2015-05-22 | 
| Date of Event | 2015-02-22 | 
| Device Manufacturer Date | 2013-10-02 | 
| Date Added to Maude | 2015-07-23 | 
| Event Key | 0 | 
| Report Source Code | Manufacturer report | 
| Manufacturer Link | Y | 
| Number of Patients in Event | 0 | 
| Adverse Event Flag | 3 | 
| Product Problem Flag | 3 | 
| Reprocessed and Reused Flag | 3 | 
| Health Professional | 3 | 
| Initial Report to FDA | 3 | 
| Report to FDA | 0 | 
| Event Location | 0 | 
| Manufacturer Contact | THOMAS TYRE | 
| Manufacturer Street | 9099 NORTH DEERBROOK TRAIL | 
| Manufacturer City | BROWN DEER WI 53223 | 
| Manufacturer Country | US | 
| Manufacturer Postal | 53223 | 
| Manufacturer G1 | INVITROGEN CORPORATION | 
| Manufacturer Street | 9099 NORTH DEERBROOK TRAIL | 
| Manufacturer City | BROWN DEER WI 53223 | 
| Manufacturer Country | US | 
| Manufacturer Postal Code | 53223 | 
| Single Use | 3 | 
| Remedial Action | RB | 
| Previous Use Code | 3 | 
| Removal Correction Number | 2244574-008-2015 C | 
| Event Type | 3 | 
| Type of Report | 3 | 
| Brand Name | ALLSET GOLD A HIGH RES SSP | 
| Generic Name | TEST, QUALITATIVE FOR HLA | 
| Product Code | MZI | 
| Date Received | 2015-07-10 | 
| Catalog Number | 54010D | 
| Lot Number | 012 1446993 | 
| Device Expiration Date | 2015-03-31 | 
| Operator | HEALTH PROFESSIONAL | 
| Device Availability | N | 
| Device Age | DA | 
| Device Eval'ed by Mfgr | R | 
| Device Sequence No | 1 | 
| Device Event Key | 0 | 
| Manufacturer | LIFE TECHNOLOGIES | 
| Manufacturer Address | BROWN DEER WI US | 
| Patient Number | Treatment | Outcome | Date | 
|---|---|---|---|
| 1 | 0 | 2015-07-10 |