B LOCUS SSP UNITRAY 20 TESTS 7870010

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,health professional report with the FDA on 2015-07-13 for B LOCUS SSP UNITRAY 20 TESTS 7870010 manufactured by Life Technologies.

Event Text Entries

[24524491] This report is related to customer complaint (b)(4) made against b locus ssp unitray 20 tests (sku 7870010, lot 027 1544274). The customer reported that the reactivity for lane 8 was negative for the b40:08 allele when the labeling indicates the reactivity should be positive. The labeling discrepancy may lead to the b40:08 allele being missing from the string of many possible alleles. The complaint was confirmed on (b)(6) 2015. The complaint and possible root cause is still under investigation.
Patient Sequence No: 1, Text Type: D, B5


[65725980] The investigation was concluded on 07/13/21015 and it has been determined to update labeling with the comment "a weak true positive at the correct size (210bp) may be observed with b*40:08 and like samples. " life technologies performed a health hazard evaluation and concluded there was a negligible health hazard. Remedial action initiated.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2244574-2015-00097
MDR Report Key4914868
Report Source05,HEALTH PROFESSIONAL
Date Received2015-07-13
Date of Report2015-06-15
Date of Event2015-06-15
Date Mfgr Received2015-06-15
Device Manufacturer Date2015-04-03
Date Added to Maude2015-08-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactTHOMAS TYRE
Manufacturer Street9099 NORTH DEERBROOK TRAIL
Manufacturer CityBROWN DEER WI 53223
Manufacturer CountryUS
Manufacturer Postal53223
Manufacturer Phone4142144025
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameB LOCUS SSP UNITRAY 20 TESTS
Product CodeMZI
Date Received2015-07-13
Catalog Number7870010
Lot Number027 1544274
Device Expiration Date2015-11-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerLIFE TECHNOLOGIES
Manufacturer AddressBROWN DEER WI US


Patients

Patient NumberTreatmentOutcomeDate
10 2015-07-13

© 2026 FDA.report