ROTATING CF RESECTOSCOPE VISUAL OBTURATOR, 25FR ERVO-CF25

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-07-17 for ROTATING CF RESECTOSCOPE VISUAL OBTURATOR, 25FR ERVO-CF25 manufactured by Gyrus Acmi Inc.

Event Text Entries

[22457372] The device referenced in this report has been returned to olympus for evaluation. The reported phenomenon which stated that the white tip of the device came off after placing in the bladder was not confirmed. A visual and mechanical testing was performed. The release button was found to be loose from the lock ring and pin assembly. The o-ring was also missing from the lock body causing the device to fail the leak test. This could be attributed to normal wear of the device. The instruction for use states, "examine the device prior to use. Do not use if damage is found. "
Patient Sequence No: 1, Text Type: N, H10


[22457373] Olympus was informed that during an unspecified procedure, the white tip of the device came off after placing in the bladder. No further information was provided. Olympus made multiple attempts to obtain additional information via telephone and in writing, but with no success.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2951238-2015-00305
MDR Report Key4922966
Date Received2015-07-17
Date of Report2015-07-02
Date of Event2015-07-02
Date Mfgr Received2015-07-02
Date Added to Maude2015-07-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS NOEMI SCHAMBACH
Manufacturer Street2400 RINGWOOD AVENUE
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4089355002
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameROTATING CF RESECTOSCOPE VISUAL OBTURATOR, 25FR
Generic NameROTATING CF RESECTOSCOPE VISUAL OBTURATOR, 25FR
Product CodeFDC
Date Received2015-07-17
Returned To Mfg2015-07-02
Model NumberERVO-CF25
Catalog NumberERVO-CF25
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerGYRUS ACMI INC
Manufacturer Address136 TURNPIKE ROAD SOUTHBOROUGH MA 017722104 US 017722104


Patients

Patient NumberTreatmentOutcomeDate
10 2015-07-17

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.