MONOMAX VIOLET 2/0 (3) 150CM HR30 (M) B0041453

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,07,08 report with the FDA on 2015-07-15 for MONOMAX VIOLET 2/0 (3) 150CM HR30 (M) B0041453 manufactured by B. Braun Surgical S.a..

Event Text Entries

[6431964] Country of complaint: (b)(6). The inner foil and outer foil are welded together, the inner package is no longer sterile when released into the sterile field.
Patient Sequence No: 1, Text Type: D, B5


[13509456] Samples received: 2 open pouches. There are no previous complaints of this code batch. There are no units in stock. We have received two open samples, only the second pack is opened. Both open samples received do not have the aluminum pouch stuck to the outer paper foil. We need defective samples showing the defect to properly assess the complaint. Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfill oem requirements. Corrective/preventive actions: not applicable.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2916714-2015-00593
MDR Report Key4926080
Report Source01,07,08
Date Received2015-07-15
Date of Report2015-07-15
Date of Event2015-06-15
Date Mfgr Received2015-06-15
Date Added to Maude2015-07-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactNICOLE BROYLES
Manufacturer Street615 LAMBERT POINTE DRIVE
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3145515988
Manufacturer G1B.BRAUN SURGICAL SA
Manufacturer Street121 CARRETERA DE TERRASSA
Manufacturer CityRUBI (BARCELONA) 08191
Manufacturer CountryES
Manufacturer Postal Code08191
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMONOMAX VIOLET 2/0 (3) 150CM HR30 (M)
Generic NameSUTURE
Product CodeNWJ
Date Received2015-07-15
Model NumberB0041453
Catalog NumberB0041453
Lot Number113463
OperatorHEALTH PROFESSIONAL
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerB. BRAUN SURGICAL S.A.
Manufacturer AddressRUBI (BARCELONA) 08191 ES 08191


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-07-15