FREE PSA 03289788190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06,foreign,health profe report with the FDA on 2015-07-23 for FREE PSA 03289788190 manufactured by Roche Diagnostics.

Event Text Entries

[6011329] The customer complained of erroneous results for 1 patient sample tested for free psa (free psa). The results were reported outside of the laboratory. The initial free psa result was 3. 79 ng/ml. The total (free + complexed) psa - prostate-specific antigen (tpsa) - (total psa) result was 0. 543 ng/ml. The sample was repeated and the free psa result was 3. 83 ng/ml. The total psa result was 0. 556 ng/ml. On (b)(6) 2015 the sample was tested on a different e601 analyzer and the free psa result was 3. 86 ng/ml. The total psa result was 0. 674 ng/ml. The sample was also tested on an architect analyzer where the free psa result was 0. 624 ng/ml. The total psa result was 1. 079 ng/ml. No adverse event occurred. The cobas e601 analyzer serial numbers were not provided.
Patient Sequence No: 1, Text Type: D, B5


[14378602] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[24989392] A specific root cause could not be identified. The patient sample could not be provided for investigation. Based on the available data, there is no indication of an issue with the system or the reagent.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2015-03858
MDR Report Key4936613
Report Source01,05,06,FOREIGN,HEALTH PROFE
Date Received2015-07-23
Date of Report2015-09-02
Date of Event2015-07-03
Date Mfgr Received2015-07-06
Date Added to Maude2015-07-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameFREE PSA
Generic NameTEST,PROSTATE SPECIFIC ANTIGEN,FREE,(NONCOMPLEXED)
Product CodeMTG
Date Received2015-07-23
Model NumberNA
Catalog Number03289788190
Lot Number183535
ID NumberNA
Device Expiration Date2016-05-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2015-07-23

© 2026 FDA.report