DEROYAL M7035-S 11342

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-07-23 for DEROYAL M7035-S 11342 manufactured by Deroyal Industires, Inc..

Event Text Entries

[22958084] Investigation findings: the complaint sample was returned and inspected by the quality control investigator. The product was well worn. The material/body was damaged from wear and use. This included wearing down of the stay pocket resulting in the stay coming out. This product, m7035-s, is a pediatric elbow positioner/immobilizer is intended to be prescribed by a properly licensed practitioner to provide comfortable support while helping prevent patients from pulling i. V. Lines, nasogastric (ng) tubes and tracheostomy tubes. It is a type of restraint and it is contraindicated for a patient who is or may become highly aggressive or agitated. Based on the type of wear is seems to have been misused as a support brace by the end user or user facility. This would also contribute to the damage seen. In the complaint text it was also recorded that the therapist suggested that the original product was too short and allowed the patient to move her arms too much. A larger size product was requested as a replacement. It specifically asked for the larger size product # m7035-l. Correction: a larger size replacement was sent. Root cause analysis: it is concluded that excess wear, misuse and, improper fitting of product when it was issued resulted in material breakdown. Corrective action and/or systemic correction action taken: no action is required at this time since no manufacturing defect was found. Preventive action: no action is required at this time since no manufacturing defect was found.
Patient Sequence No: 1, Text Type: N, H10


[22958085] This complaint is on product number m7035-s. This is a pediatric elbow positioner/immobilizer is intended to be prescribed by a properly licensed practitioner to provide comfortable support while helping prevent patients from pulling i. V. Lines, nasogastric (ng) tubes and tracheostomy tubes. It is a type of restraint and it is contraindicated for a patient who is or may become highly aggressive or agitated. The customer reported that "metal stays have come out of material pockets and patient has cuts on her forearm and biceps. " the patient had cuts from the stays were treated with topical ointment.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3010452421-2015-00009
MDR Report Key4936809
Date Received2015-07-23
Date of Report2015-07-22
Date of Event2015-07-01
Date Facility Aware2015-07-01
Report Date2015-07-01
Date Reported to Mfgr2015-07-01
Date Mfgr Received2015-07-01
Date Added to Maude2015-07-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDR. MARIAN VARGAS
Manufacturer Street200 DEBUSK LN
Manufacturer CityPOWELL TN 37849
Manufacturer CountryUS
Manufacturer Postal37849
Manufacturer Phone8653621013
Manufacturer G1DEROYAL INDUSTRIES, INC.
Manufacturer StreetKM 20.5 CARRETERA A VILLA CANA
Manufacturer CityVILLA CANALES, 01065
Manufacturer CountryGT
Manufacturer Postal Code01065
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameDEROYAL
Generic NameELBOW POSITIONER AND IMMOBILIZER WITH METAL STAYS
Product CodeILI
Date Received2015-07-23
Returned To Mfg2015-07-14
Model NumberM7035-S
Catalog Number11342
Lot Number39173312
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerDEROYAL INDUSTIRES, INC.
Manufacturer Address200 DEBUSK LANE POWELL TN 37849 US 37849


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-07-23

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.