FREE PSA 03289788160

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2015-07-27 for FREE PSA 03289788160 manufactured by Roche Diagnostics.

Event Text Entries

[6037745] The customer received questionable total psa total (free + complexed) psa - prostate-specific antigen (total psa) and free psa results for one patient sample. Refer to the medwatch with patient identifier (b)(6) for the total psa assay. The initial total psa result from the cobas 8000 analyzer was 0. 750 ng/ml. The repeat result from a cobas 6000 analyzer was 0. 753 ng/ml. The initial free psa result from cobas e411 serial number (b)(4) was 3. 43 ng/ml. The repeat result was 3. 44 ng/ml. The total psa result of 0. 750 ng/ml and free psa result of 3. 43 ng/ml were reported outside the laboratory. The patient was not adversely affected. As the sample was no longer available for investigation, a specific root cause could not be determined. A general reagent issue was not suspected.
Patient Sequence No: 1, Text Type: D, B5


[13712553] This event occurred in (b)(6). (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2015-03864
MDR Report Key4943003
Report Source01,05,06
Date Received2015-07-27
Date of Report2015-07-27
Date of Event2015-07-09
Date Mfgr Received2015-07-09
Date Added to Maude2015-07-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameFREE PSA
Generic NameTEST,PROSTATE SPECIFIC ANTIGEN,FREE,(NONCOMPLEXED) TO DISTINGUISH PROSTATE CA
Product CodeMTG
Date Received2015-07-27
Model NumberNA
Catalog Number03289788160
Lot Number178872
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2015-07-27

© 2026 FDA.report