[6256282] At 0144 in the morning, a lithium-ion battery used to power a hospital workstation on wheels caught fire while in its wall mounted charging station. A hospital staff member heard hissing and popping, followed by smoke then fire; 911 was called and the hospital's intensive care pts had to be evacuated. Five hospital staff members reported inhaling smoke, but no injuries otherwise resulted. Clean-up on the unit took a week to complete. This is the second time this year a battery from the same company has resulted in a fire within our hospital system. While the battery is not a medical device per se, the malfunction affected pt care. An ecri alert was released related to the matter.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5044781
MDR Report Key4945812
Date Received2015-07-21
Date of Report2015-07-21
Date of Event2015-06-04
Date Added to Maude2015-07-28
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameMOBIUS POWER MEDICAL CART BATTERY
Generic NameLITHIUM-ION BATTERY FOR CART
Product CodeFCO
Date Received2015-07-21
Returned To Mfg2015-06-05
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Sequence No1
Device Event Key0
ManufacturerENOVATE MEDICAL
Manufacturer AddressMURFREESBORO TN US


Patients

Patient NumberTreatmentOutcomeDate
101. Other; 2. Required No Informationntervention 2015-07-21