FASTLOAD CTA 6730 017347

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-07-23 for FASTLOAD CTA 6730 017347 manufactured by Bracco Injennering Sa.

Event Text Entries

[6633926] Product defect in manufacturing of the fastload cta dual syringe pack used in the empower cta and empower cta injector systems. Two of three packages had the locking lugs on the syringe broken off. One of three had the low pressure connecting y-tube blue cap crushed and torn off. These all occurred in the same lot number. None were used on any patients, however, the broken locking lugs on the syringes made them unusable. None of the other dual packs in stock with this lot number had the same defect. Attempts were made to duplicate the error by dropping the packages from 5 feet to the floor. There was no damage. Fastload cta dual syringe pack with spikes contains: 2x 200ml fastload syringes, 1 low pressure connecting t tube, 1 small spike and 1 large spike.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5044858
MDR Report Key4949666
Date Received2015-07-23
Date of Report2015-07-23
Date of Event2015-07-22
Date Added to Maude2015-07-30
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameFASTLOAD CTA
Generic NameINJECTOR SYSTEM
Product CodeDXT
Date Received2015-07-23
Model Number6730
Catalog Number017347
Lot NumberIADA4-1505
Device Expiration Date2018-05-31
OperatorHEALTH PROFESSIONAL
Device Sequence No1
Device Event Key0
ManufacturerBRACCO INJENNERING SA
Manufacturer AddressLAUSANNE SZ

Device Sequence Number: 2

Brand NameFASTLOAD CTA
Generic NameINJECTOR SYSTEM
Product CodeIZQ
Date Received2015-07-23
Model Number6730
Catalog Number017347
Lot NumberIADA4-1505
Device Sequence No2
Device Event Key0
ManufacturerBRACCO INJENEERING SA
Manufacturer AddressLUSANNE, CH SZ

Device Sequence Number: 3

Brand NameFASTLOAD CTA
Generic NameINJECTOR SYSTEM
Product CodeIZQ
Date Received2015-07-23
Model Number6730
Catalog Number017347
Lot NumberIADA4-1505
Device Expiration Date2018-05-31
Device Sequence No3
Device Event Key0
ManufacturerBRACCO INJENEERING SA
Manufacturer AddressLAUSANNE, CH SZ


Patients

Patient NumberTreatmentOutcomeDate
10 2015-07-23

© 2026 FDA.report