BLUE COMFORT BOUFFANT 24 IN 3274

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-07-29 for BLUE COMFORT BOUFFANT 24 IN 3274 manufactured by Cardinal Health.

Event Text Entries

[23155288] Multiple attempts to obtain samples from the customer were made and the samples have just been received. Therefore; a follow up report will be filed once the sample evaluation has been completed.
Patient Sequence No: 1, Text Type: N, H10


[23155289] The patient has allergy conditions and developed a rash around her hairline where the cap fit. Caps are latex free.
Patient Sequence No: 1, Text Type: D, B5


[34205919] The customer's sample was received for evaluation. A review of the returned sample did not indicate any abnormal findings that would result in the allergic reaction. A review of the device history record could not be conducted as no lot number was provided. A review of raw material records for the current production indicated that the material used was approved and in compliance with the required specifications. The material used for the bouffant is polypropylene with calcium carbonate fillers, which passed the biocompatibility tests prescribed by the regulatory agency for the intended use. There were no changes to the material composition including elastic and thread. Therefore, based on these findings we are unable to determine the root cause for the issue reported. No corrective action will be taken at this time. We will continue to monitor for any similar reports.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1423537-2015-00055
MDR Report Key4950675
Date Received2015-07-29
Date of Report2015-07-29
Date of Event2015-06-29
Date Mfgr Received2015-06-30
Date Added to Maude2015-07-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMICHELE DONATICH
Manufacturer Street1500 WAUKEGAN ROAD
Manufacturer CityWAUKEGAN IL 60085
Manufacturer CountryUS
Manufacturer Postal60085
Manufacturer Phone8478876412
Manufacturer G1CARDINAL HEALTH SINGAPORE 225 PTE LTD
Manufacturer Street10 KALLANG AVENUE #10-10/18 APERIA
Manufacturer CitySINGAPORE 339510
Manufacturer CountrySN
Manufacturer Postal Code339510
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBLUE COMFORT BOUFFANT 24 IN
Generic NameBLUE COMFORT BOUFFANT
Product CodeFYF
Date Received2015-07-29
Returned To Mfg2015-07-27
Model Number3274
Catalog Number3274
Lot NumberNO LOT GIVEN
OperatorLAY USER/PATIENT
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCARDINAL HEALTH
Manufacturer Address1500 WAUKEGAN ROAD WAUKEGAN IL 60085 US 60085


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-07-29

© 2026 FDA.report