Event Text Entries

[6035331] The customer received questionable total (free + complexed) psa - prostate-specific antigen (tpsa) and free psa results for multiple samples from one patient. The samples were tested on cobas e601 serial number (b)(4). Refer to the medwatch with patient identifier (b)(6) for the total psa assay. On 05/27/2015, the tpsa was 0. 32 ng/ml and free psa was 0. 42 ng/ml. On (b)(6) 2015, the tpsa was 0. 235 ng/ml and free psa was 0. 513 ng/ml. The sample was repeated and the tpsa was 0. 232 ng/ml and free psa was 0. 516 ng/ml. The sample was tested on cobas e601 serial number (b)(4) and the tpsa was 0. 219 ng/ml and free psa was 0. 354 ng/ml. A second sample drawn (b)(6) 2015 was tested and the tpsa was 0. 227 ng/ml and free psa was 0. 508 ng/ml. The erroneous results were reported outside the laboratory. The patient was not adversely affected. The free psa reagent expiration date was given as "2016-05".
Patient Sequence No: 1, Text Type: D, B5


[13967573] This event occurred in (b)(6). Unique identifier (udi)#: asku.
Patient Sequence No: 1, Text Type: N, H10


[26785702] Sample from the patient was submitted for investigation and the customer's results were confirmed on a cobas (b)(4). No anti-idiotype or (b)(4) interferences were found in the sample. An unknown interfering substance was identified by adding a psa specific antibody with different specificity to the tpsa reagent. This type of interference is described in product labeling. Further investigation was not possible as no sample material remained.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2015-03887
MDR Report Key4953349
Report Source01,05,06,FOREIGN,HEALTH PROFE
Date Received2015-07-30
Date of Report2015-09-02
Date of Event2015-05-27
Date Mfgr Received2015-07-14
Date Added to Maude2015-07-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameFREE PSA
Generic NameTEST,PROSTATE SPECIFIC ANTIGEN,FREE,(NONCOMPLEXED) TO DISTINGUISH PROSTATE CA
Product CodeMTG
Date Received2015-07-30
Model NumberNA
Catalog NumberASKU
Lot Number183535
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250