CHROMID SALMONELLA AGAR 43621

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-07-31 for CHROMID SALMONELLA AGAR 43621 manufactured by Biom?rieux Sa.

Event Text Entries

[24654713] Device not returned to manufacturer.
Patient Sequence No: 1, Text Type: N, H10


[24654714] A clinical customer in (b)(6) contacted biomerieux to report the observation of white colonies on chromid[tm] salmonella agar (ref. (b)(4), lot 1003998840) when the customer was expecting to see the typical pink/mauve colonies indicative of a positive sample. The customer classified this as a false negative result. The negative result was reported to the physician. There is no indication or evidence of patient death or serious injury as a result of the event. The referenced product is for sale in the united states for industry use only, and is not used to test patient samples; however, were the event to recur while testing products for human consumption, contaminated product could be introduced into market. Additionally, similar chromid agar products are sold in the united states for clinical use. Biomerieux has requested the patient isolate for internal investigation.
Patient Sequence No: 1, Text Type: D, B5


[34346806] An investigation was conducted using the chromid tm salmonella retained samples of the customer batch (1003998840) in comparison with a random batch of chromid salmonella as reference. Sample strains were the same as those used during release tests. Good growth was observed for all quality control salmonella strains on the customer batch (1003998840) with characteristic pale pink-mauve colonies after 24h of incubation, and dark mauve colonies after 48h of incubation. The same results were observed for the reference batch. In addition, testing was performed with the strain submitted by the customer on the customer batch (1003998840) and on a random batch of chromid salmonella as reference. The same results were observed for the two batches. Good growth was observed for the customer's strain, with uncolored colonies after 24 hours of incubation at 33-37? C and pale mauve colonies after 48 hours of incubation. It was observed that the customer's strain presents a slow growth whatever the chromid salmonella batch used, and therefore the reported issue is not related to any defect with the performance of the chromid salmonella agar ref 43621, lot 1003998840. As indicated in the package insert of this medium, some salmonella serovars may produce a weak coloration or be slow to produce a coloration. The review of complaint records indicates that no other complaint has been registered against this batch number. During the review of the manufacturing and qc batch records, we have identified the following information: this batch (number 1003998840, expiry date: 2015-09-30) was manufactured 13may2015. No discrepancies were detected during the manufacturing of this batch. For the release of the batch after its manufacturing, the results of the microbiological quality control are in accordance with specifications as indicated in the quality certificate.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002769706-2015-00062
MDR Report Key4957213
Date Received2015-07-31
Date of Report2015-06-29
Date Mfgr Received2015-06-29
Device Manufacturer Date2015-05-13
Date Added to Maude2015-07-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. RYAN LEMELLE
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147318582
Manufacturer G1BIOMERIEUX, INC
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityST. LOUIS MO 63042
Manufacturer CountryUS
Manufacturer Postal Code63042
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHROMID SALMONELLA AGAR
Generic NameCHROMID SALMONELLA AGAR
Product CodeJWX
Date Received2015-07-31
Catalog Number43621
Lot Number1003998840
Device Expiration Date2015-09-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOM?RIEUX SA
Manufacturer Address5, RUE DES AQUEDUCS CRAPONNE, FR 69290 FR 69290


Patients

Patient NumberTreatmentOutcomeDate
10 2015-07-31

© 2026 FDA.report