CLINIMACS CD34 REAGENT 171-01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-07-22 for CLINIMACS CD34 REAGENT 171-01 manufactured by Miltenyi Biotec Gmbh.

Event Text Entries

[24526044] Conclusion: this complaint is classified as application failure.
Patient Sequence No: 1, Text Type: N, H10


[24526045] The customer reported that one of their clinimacs cd34 reagents failed in regard that the grey rubber stopper in the vial completely collapsed in on itself when inserting the universal vial adapter (mckesson, cat #ur12b20109) in order to remove the reagent and transfer it to a bag. This resulted in the cd34 reagent being exposed to the environment and not sterile, rendering it unacceptable for use in the cell manufacturing procedure. There was a backup vial of the clinimacs cd34 reagent in place which was used to successfully complete the cell separation procedure. Therefore the patient's treatment was not compromised. Based on the information provided an adverse event for the patient can be excluded. The event happened at: (b)(6).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005290010-2015-00006
MDR Report Key4959955
Date Received2015-07-22
Date of Report2015-07-21
Date of Event2015-06-17
Date Mfgr Received2015-06-28
Date Added to Maude2015-08-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNANCY JOHANSEN
Manufacturer Street85 HAMILTON ST.
Manufacturer CityCAMBRIDGE MA 02139
Manufacturer CountryUS
Manufacturer Postal02139
Manufacturer Phone6172180062
Manufacturer G1MILTENYI BIOTEC GMBH
Manufacturer StreetFRIEDRICH-EBERT-STRASSE 68
Manufacturer City51429
Manufacturer CountryGM
Manufacturer Postal Code51429
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCLINIMACS CD34 REAGENT
Generic NameCLINIMACS CD34 REAGENT
Product CodeKSR
Date Received2015-07-22
Model NumberNA
Catalog Number171-01
Lot Number6150407007
ID Number04049934004827
Device Expiration Date2015-09-07
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMILTENYI BIOTEC GMBH
Manufacturer AddressFRIEDRICH-EBERT-STRASSE 68 BERGISCH 1 GLADBACH 51429 GM 51429


Patients

Patient NumberTreatmentOutcomeDate
10 2015-07-22