TI500 GLOBETROTTER *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00 report with the FDA on 2003-11-24 for TI500 GLOBETROTTER * manufactured by Hill-rom Airshields.

Event Text Entries

[295416] Incident was filed in regards to this globetrotter unit. He is reporting that during the transport of a pt the flow of the o2 stopped during the transport and the pt required immediate care when they arrived at the icu. No injury is being reported. The customer is alleging that they have replaced the flow valve p/n 6736249 several times due to problems with the knob binding. 2003 tsr on site. Found that starting from off and selecting low flows up to 6 lpm, there is no flow. At 6 lpm the flowmeter starts delivering gas. When lower down to 1/2 lpm the flow control seems to function. This is the 3rd time the flow control has been replaced.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2510954-2003-00008
MDR Report Key498623
Report Source00
Date Received2003-11-24
Date of Report2003-11-21
Date of Event2003-10-21
Date Mfgr Received2003-10-21
Device Manufacturer Date2001-11-01
Date Added to Maude2003-12-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactMONICA FERRANTE
Manufacturer Street330 JACKSONVILLE RD
Manufacturer CityHATBORO PA 19040
Manufacturer CountryUS
Manufacturer Postal19040
Manufacturer Phone2156828691
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTI500 GLOBETROTTER
Generic NameTRANSPORT INCUBATOR
Product CodeFPL
Date Received2003-11-24
Model NumberTI500
Catalog Number*
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key487371
ManufacturerHILL-ROM AIRSHIELDS
Manufacturer Address330 JACKSONVILLE RD. HATBORO PA 19040 US
Baseline Brand NameTI500 GLOBETROTTER
Baseline Generic NameTRANSPORT INCUBATOR
Baseline Model NoTI500
Baseline Catalog No*
Baseline ID*


Patients

Patient NumberTreatmentOutcomeDate
10 2003-11-24

© 2026 FDA.report