CASCADE TRANSFER LIFT 383000-1L

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2015-08-03 for CASCADE TRANSFER LIFT 383000-1L manufactured by Penner Manufacturing, Inc..

Event Text Entries

[23965959] Device was inspected on 04/24/2015 and found to be in good condition. Interviews were conducted along with in house training on lift transfer. Employees were trained to use the brakes and all supplied belts during use with residents. Prior to inspection additional safety belts were ordered and rec'd by facility. Both the maintenance supervisor and the director of nursing both said that they believed that the belts were not being used at the time of the incident. It was also discussed that because of the size of the resident that more than one cna should be helping because one isn't enough to help just due to the size of a (b)(6) pound resident and the cna who is just over 5' and just over 100 pounds.
Patient Sequence No: 1, Text Type: N, H10


[23965960] Resident had just been given a bath and was taken out of the tub by the cna and resident went to reach for a towel to help cna dry him off and resident fell forward out of transfer and hit head on floor of bathing room. Resident was taken to the emergency room and evaluated. It was discovered that the resident had a concussion and a black eye.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1922538-2015-00001
MDR Report Key4990447
Report SourceCOMPANY REPRESENTATIVE
Date Received2015-08-03
Date of Report2015-08-03
Date of Event2015-04-17
Date Facility Aware2015-04-20
Report Date2015-08-03
Date Mfgr Received2015-04-20
Device Manufacturer Date2013-06-01
Date Added to Maude2015-08-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPATRICK WALL
Manufacturer Street101 GRANT ST
Manufacturer CityAURORA NE 68818
Manufacturer CountryUS
Manufacturer Postal68818
Manufacturer Phone4026945003
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCASCADE TRANSFER LIFT
Generic NameCASCADE TRANSFER LIFT
Product CodeFNG
Date Received2015-08-03
Model Number383000-1L
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age2 YR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerPENNER MANUFACTURING, INC.


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-08-03

© 2026 FDA.report