HYDRA IRRIGATION SYSTEM SIT-355 SIT-355-10

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2015-08-14 for HYDRA IRRIGATION SYSTEM SIT-355 SIT-355-10 manufactured by Endochoice Inc..

Event Text Entries

[24616534] The hydra irrigation tubing set was in use with a medivators 24-hour connector, which does not include a check valve. Therefore the only back flow valve was on the hydra irrigation tubing set. It was determined upon testing of the returned device that the valve did not function properly. Subsequent testing of this lot found no other units exhibiting this issue. Endochoice has recommended to the site that they utilize single-use connectors that include back flow valves.
Patient Sequence No: 1, Text Type: N, H10


[24616535] A user reported that liquid gastric contents back-flowed into the irrigation tubing during a gastroscopy. There was no patient impact. An olympus 190 gastroscope was in use with mediators endogator 24-hour use irrigation port connector and the endochoice hydra irrigation tubing set. Submission of this report does not, in itself, represent a conclusion that medical device caused or contributed to an adverse event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007591333-2015-00035
MDR Report Key5008543
Report SourceCOMPANY REPRESENTATIVE
Date Received2015-08-14
Date of Report2015-08-14
Date of Event2015-07-22
Date Mfgr Received2015-07-28
Date Added to Maude2015-08-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street11810 WILLS RD
Manufacturer CityALPHARETTA GA 30009
Manufacturer CountryUS
Manufacturer Postal30009
Manufacturer Phone6787084743
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameHYDRA IRRIGATION SYSTEM
Generic NamePUMP, AIR, NON-MANUAL, FOR ENDOSCOPE
Product CodeFEQ
Date Received2015-08-14
Returned To Mfg2015-07-27
Model NumberSIT-355
Catalog NumberSIT-355-10
Lot Number140520
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerENDOCHOICE INC.
Manufacturer Address11810 WILLS RD ALPHARETTA GA 300090000 US 300090000


Patients

Patient NumberTreatmentOutcomeDate
10 2015-08-14

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.