SUPPORT ARM 177 6481720

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2015-08-13 for SUPPORT ARM 177 6481720 manufactured by Maquet Critical Care Ab.

Event Text Entries

[24050794] Further info regarding the event has been sought. A supplemental medwatch report will be submitted when the investigation is completed.
Patient Sequence No: 1, Text Type: N, H10


[24050795] It was reported that the support arm for the ventilator broke. There was no patient involvement. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[42201683] (b)(4). According to information received, the support arm broke at the joint above the first bend. The support arm is by design a casting product. No parts or pictures were received back for investigation. Neither was information of how or under which circumstances the support arm broke received. Previous investigations of the support arm led to a change of the manufacturing process in order to obtain a higher mechanical strength of the support arm. This change was implemented in production during september 2009. It is unknown if the reported support arm was manufactured before, or after, the implementation of this change. Therefore, the cause for this event cannot be determined.
Patient Sequence No: 1, Text Type: N, H10


[42201684] (b)(4). This follow-up medwatch was created from a previously received initial medwatch with manufacturer report # : 8010042-2015-00356. The original date of the initial medwatch report was 07/24/2015. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8010042-2015-00356
MDR Report Key5015125
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2015-08-13
Date of Report2015-07-24
Date Mfgr Received2015-07-24
Date Added to Maude2015-08-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMAGNUS LINDQVIST
Manufacturer StreetMAQUET CRITICAL CARE AB R
Manufacturer CitySOLNA
Manufacturer CountrySW
Manufacturer Phone7307705
Manufacturer G1MAGNUS LINDQVIST
Manufacturer StreetMAQUET CRITICAL CARE AB R
Manufacturer CitySOLNA
Manufacturer CountrySW
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSUPPORT ARM 177
Generic NameSUPPORT, ARM
Product CodeIOY
Date Received2015-08-13
Catalog Number6481720
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerMAQUET CRITICAL CARE AB
Manufacturer AddressSOLNA SW


Patients

Patient NumberTreatmentOutcomeDate
10 2015-08-13

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.