RUSCH MRI COND FIBER OPTIC LARYN KIT 005852300

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2015-08-19 for RUSCH MRI COND FIBER OPTIC LARYN KIT 005852300 manufactured by Teleflex Medical.

Event Text Entries

[23527056] (b)(4). Upon receipt the gold pipe light and bulb base were detached from the blade. No obvious damage noted visually. The gold pipe light was re-installed onto the blade with no fitting or alignment issues noted. The blade was attached to a known good handle and fitting/alignment was not an issue. It was noted that the light was not burning after the blade was engaged. After further investigation it was determined the small incandescent bulb was burned out. The original complaint of the pipe light not fitting onto the blade cannot be confirmed. The complaint cannot be confirmed. Root cause cannot be established. No corrective/preventative actions will be assigned. No further action required.
Patient Sequence No: 1, Text Type: N, H10


[23527057] The customer alleges having difficulty reassembling the gold pipe light onto the blade. The patient's condition is reported as fine.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1044475-2015-00311
MDR Report Key5015166
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2015-08-19
Date of Report2015-07-30
Date of Event2015-07-29
Date Mfgr Received2015-07-30
Date Added to Maude2015-08-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMARGIE BURTON, RN
Manufacturer Street3015 CARRINGTON MILL BLVD
Manufacturer CityMORRISVILLE NC 27560
Manufacturer CountryUS
Manufacturer Postal27560
Manufacturer Phone9194334965
Manufacturer G1TELEFLEX MEDICAL
Manufacturer Street2917 WECK DRIVE
Manufacturer CityRESEARCH TRIANGLE PARK NC 27709
Manufacturer CountryUS
Manufacturer Postal Code27709
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRUSCH MRI COND FIBER OPTIC LARYN KIT
Generic NameLARYNGOSCOPE KIT
Product CodeOGH
Date Received2015-08-19
Returned To Mfg2015-08-11
Catalog Number005852300
Lot NumberUNKNOWN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTELEFLEX MEDICAL
Manufacturer AddressRESEARCH TRIANGLE PARK NC


Patients

Patient NumberTreatmentOutcomeDate
10 2015-08-19

© 2026 FDA.report