ADVANTA FRAME 1600

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2015-08-20 for ADVANTA FRAME 1600 manufactured by Hill-rom Batesville.

Event Text Entries

[23659337] Hill-rom technical support suggested to check the side rail switches. Per the hill-rom service manual the advanta bed requires an effective maintenance program. Preventative maintenance will minimize downtime due to excessive wear. Run the head section to the full upper and lower limits to ensure proper function of the limit switches. A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed. It is unknown if the facility performs preventative maintenance on their beds. Three attempts have been made regarding a resolution to this contact line, with no response. Based on this information, no further action is required.
Patient Sequence No: 1, Text Type: N, H10


[23659338] Hill-rom received a report from the account stating the head section would self-run when the bed is plugged in. The bed was located in a regular patient care room at the account. There was no patient/user injury reported. This report was filed in our complaint handling system as complaint #(b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1824206-2015-00822
MDR Report Key5017795
Report SourceUSER FACILITY
Date Received2015-08-20
Date of Report2015-07-24
Date of Event2015-07-24
Date Mfgr Received2015-07-24
Device Manufacturer Date2003-01-08
Date Added to Maude2015-08-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactTONY WERNER
Manufacturer Street1069 STATE ROUTE 46 EAST
Manufacturer CityBATESVILLE IN 47006
Manufacturer CountryUS
Manufacturer Postal47006
Manufacturer Phone8129312359
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameADVANTA FRAME
Generic NameA/C POWERED ADJUSTABLE HOSPITAL BED
Product CodeFNK
Date Received2015-08-20
Model Number1600
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerHILL-ROM BATESVILLE
Manufacturer Address1069 STATE ROUTE 46 EAST BATESVILLE IN 47006 US 47006


Patients

Patient NumberTreatmentOutcomeDate
10 2015-08-20

© 2024 FDA.report