PROGRESSA FRAME P7500A000180

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2015-08-20 for PROGRESSA FRAME P7500A000180 manufactured by Hill-rom Batesville.

Event Text Entries

[23659371] Hill-rom technical support suggested to switch the communication cable. It is necessary for the progressa? Bed to have an effective maintenance program. We recommend that you do annual preventive maintenance (examine and test the communication junction box. Make sure the sidecom? Communication system feature works correctly. Examine the communication cable, include the male and female pins in the plug. A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed. It is unknown if the facility performs preventative maintenance on their beds. The account replaced the communication cable to resolve the issue. Based on this information, no further action is required.
Patient Sequence No: 1, Text Type: N, H10


[23659372] Hill-rom received a report from the account stating the nurse call did not work. The bed was located in icu at the account. There was no patient/user injury reported. This report was filed in our complaint handling system as complaint #(b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1824206-2015-00824
MDR Report Key5017851
Report SourceUSER FACILITY
Date Received2015-08-20
Date of Report2015-07-27
Date of Event2015-07-27
Date Mfgr Received2015-07-27
Device Manufacturer Date2015-03-09
Date Added to Maude2015-08-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactTONY WERNER
Manufacturer Street1069 STATE ROUTE 46 EAST
Manufacturer CityBATESVILLE IN 47006
Manufacturer CountryUS
Manufacturer Postal47006
Manufacturer Phone8129312359
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePROGRESSA FRAME
Generic NameA/C POWERED ADJUSTABLE HOSPITAL BED
Product CodeFNK
Date Received2015-08-20
Model NumberP7500A000180
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerHILL-ROM BATESVILLE
Manufacturer Address1069 STATE ROUTE 46 EAST BATESVILLE IN 47006 US 47006