CAREASSIST BED CA100 1170

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-08-20 for CAREASSIST BED CA100 1170 manufactured by Hill-rom France.

Event Text Entries

[23862816] The hill-rom technician found the brakes were not holding due to three brake casters, one steer caster, and four caster supports damaged. Per the hill-rom service manual the bed should be subject to an effective maintenance program. An annual service of the bed is advised in order to maintain its characteristics and performance. Brake casters should be checked for cuts, wear and quality of tread, etc. And replaced when necessary. Check the brakes to see whether the bed moves when the brake pedals are pressed and repair as necessary. A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in 2012-2015. It is unknown if the facility performed any other preventative maintenance on this bed. The technician replaced the brake casters, one steer caster, and four caster supports to resolve the issue. Based on this information, no further action is required.
Patient Sequence No: 1, Text Type: N, H10


[23862817] Hill-rom received a report from the account stating the brakes were not holding. The bed was located at the account. There was no patient/user injury reported. This report was filed in our complaint handling system as complaint #(b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9615739-2015-00250
MDR Report Key5018309
Date Received2015-08-20
Date of Report2015-07-30
Date of Event2015-07-30
Date Mfgr Received2015-07-30
Date Added to Maude2015-08-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJOHN CUMMINGS
Manufacturer Street1069 STATE ROUTE 46 EAST
Manufacturer CityBATESVILLE IN 47006
Manufacturer CountryUS
Manufacturer Postal47006
Manufacturer Phone8129312869
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCAREASSIST BED CA100
Generic NameA/C POWERED ADJUSTABLE HOSPITAL BED
Product CodeFNK
Date Received2015-08-20
Model Number1170
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHILL-ROM FRANCE
Manufacturer Address188 RUE DU CADUCEE-PARC , EUROMEDECINE MONTPELLIER, HERAULT 34195 FR 34195


Patients

Patient NumberTreatmentOutcomeDate
10 2015-08-20

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.