VITROS CHEMISTRY PRODUCTS AMON SLIDES 1726926

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-08-20 for VITROS CHEMISTRY PRODUCTS AMON SLIDES 1726926 manufactured by Ortho-clinical Diagnostics.

Event Text Entries

[25143973] The investigation determined that an unexpected, nonreproducible vitros amon result was obtained from a single patient sample processed using vitros amon slides on a vitros 250 chemistry system. The assignable cause for the event could not be determined with the information available, however the possibility that an unexpected vitros amon reagent performance, pre-analytical handling issue or an unexpected vitros 250 system issue had contributed to the result could not be ruled out.
Patient Sequence No: 1, Text Type: N, H10


[25143974] The customer complained that an unexpected, nonreproducible vitros amon result was obtained from a single patient sample processed using vitros amon slides on a vitros 250 chemistry system. Vitros amon result: 132 vs expected 88 umol/l. Biased results of the direction and magnitude observed may lead to inappropriate physician action. The technical solution center was unable to determine what amon result, if any, was reported out of the laboratory for the affected patient. There was no allegation of patient harm as a result of this event. This report corresponds to ortho clinical diagnostics inc. Complaint number (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1319809-2015-00101
MDR Report Key5019464
Date Received2015-08-20
Date of Report2015-08-20
Date of Event2015-07-28
Date Mfgr Received2015-07-24
Device Manufacturer Date2014-09-28
Date Added to Maude2015-08-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JAMES A STEVENS
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal14626
Manufacturer Phone5854533000
Manufacturer G1ORTHO-CLINICAL DIAGNOSTICS
Manufacturer Street513 TECHNOLOGY BLVD.
Manufacturer CityROCHESTER NY 14652
Manufacturer CountryUS
Manufacturer Postal Code14652
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameVITROS CHEMISTRY PRODUCTS AMON SLIDES
Generic NameIN-VITRO DIAGNOSTIC
Product CodeJID
Date Received2015-08-20
Catalog Number1726926
Lot Number1015-0238-7392
Device Expiration Date2016-03-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerORTHO-CLINICAL DIAGNOSTICS
Manufacturer Address100 INDIGO CREEK DRIVE ROCHESTER NY 14626 US 14626


Patients

Patient NumberTreatmentOutcomeDate
10 2015-08-20