VERSACARE FRAME P3200K000472

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2015-08-21 for VERSACARE FRAME P3200K000472 manufactured by Hill-rom Batesville.

Event Text Entries

[23720266] The hill-rom technician found the external alarm wire was not fully connected. Per the hill-rom service manual, perform annual preventive maintenance procedures to make sure all versacare? Bed components are functioning as originally designed. Brake casters should be checked for cuts, wear and quality of tread, etc. And replaced when necessary. Examine the brakes to see whether the bed moves when the brake are set. Replace as necessary. A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed. It is unknown if the facility performs preventative maintenance on their beds. The technician reconnected the external alarm wire to resolve the issue. Based on this information, no further action is required.
Patient Sequence No: 1, Text Type: N, H10


[23720267] Hill-rom received a report from a hill-rom technician stating the brakes not set alarm did not work. The bed was located in the warehouse at the account. There was no patient/user injury reported. This report was filed in our complaint handling system as (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1824206-2015-00834
MDR Report Key5021105
Report SourceCOMPANY REPRESENTATIVE
Date Received2015-08-21
Date of Report2015-07-30
Date of Event2015-07-30
Date Mfgr Received2015-07-30
Device Manufacturer Date2015-07-14
Date Added to Maude2015-08-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactTONY WERNER
Manufacturer Street1069 STATE ROUTE 46 EAST
Manufacturer CityBATESVILLE IN 47006
Manufacturer CountryUS
Manufacturer Postal47006
Manufacturer Phone8129312359
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVERSACARE FRAME
Generic NameA/C POWERED ADJUSTABLE HOSPITAL BED
Product CodeFNK
Date Received2015-08-21
Model NumberP3200K000472
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHILL-ROM BATESVILLE
Manufacturer Address1069 STATE ROUTE 46 EAST BATESVILLE IN 47006 US 47006


Patients

Patient NumberTreatmentOutcomeDate
10 2015-08-21

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.