CYSTOURETHROSCOPE SHEATHS 23FR E123-O

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-08-21 for CYSTOURETHROSCOPE SHEATHS 23FR E123-O manufactured by Gyrus Acmi Inc,.

Event Text Entries

[24679792] The device referenced in this report was returned to olympus for evaluation. The evaluation could not confirm the reported event. A visual inspection of the device found no scrapes, dents or other damage that shows metal or pieces of the device fell off the unit. The exact cause of the reported complaint could not be conclusively determined. Cross reference: 2951238-2015-00395, 2951238-2015-00400.
Patient Sequence No: 1, Text Type: N, H10


[24679793] Olympus was informed that during a diagnostic total laparoscopic hysterectomy procedure, metal flakes from several different components on the device were falling off into the patient's bladder. The surgeon used a grasper and successfully retrieved all the device fragments. The intended procedure was completed with the same device. There was no patient injury reported. The patient was discharged home with no complications. Olympus followed up with the user facility to obtain additional information regarding the reported event and was informed that the device was inspected and tested prior to the procedure with no anomalies found. This is two of three reports.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2951238-2015-00399
MDR Report Key5021504
Date Received2015-08-21
Date of Report2015-08-06
Date of Event2015-08-06
Date Mfgr Received2015-08-06
Date Added to Maude2015-08-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. NOEMI SCHAMBACH
Manufacturer Street2400 RINGWOOD AVENUE
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4089355002
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCYSTOURETHROSCOPE SHEATHS 23FR
Generic NameSHEATH WITH STANDARD OBTURATOR
Product CodeFEC
Date Received2015-08-21
Returned To Mfg2015-08-12
Model NumberE123-O
Catalog NumberE123-O
Lot NumberGP
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerGYRUS ACMI INC,
Manufacturer Address136 TURNPIKE ROAD SOUTHBOROUGH MA 018822104 US 018822104


Patients

Patient NumberTreatmentOutcomeDate
10 2015-08-21

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.