ROHO HIGH PROFILE CUSHION

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2015-08-11 for ROHO HIGH PROFILE CUSHION manufactured by Roho, Inc..

Event Text Entries

[23924315] Roho was not made aware of this injury until approx. 4 years after it took place. The nurse that contacted roho is working for an attorney and has not personally evaluated the end user. She did not give roho any patient identifying information due to hippa, therefore, roho does not have any means of contacting the end user or evaluating the cushion that was in use at the time of the car accident. Based on the information that was provided by the nurse, there may have been several different factors contributing to the end user's injury, including but not limited to improper inflation of her cushion, extended periods of sitting under pressure relief, and a car accident. If additional information is obtained, a supplement report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[23924316] Roho was contacted on (b)(6) 2015 by a nurse consultant working for an attorney. All of our information on the patient and her injury came from two phone conversations with this nurse. The nurse stated the following during the initial call: the patient is a c5-c6 quadriplegic. The patient was driving her car and got hit on the right side, then hit a tree. The impact was at 25-30 mph. The patient claims that the valve on her roho cushion broke and her cushion went flat. There is no information on how the valve was broken or if any action was taken to repair it or to get a replacement cushion. The patient developed a wound that she blames on the flat cushion. She was later hospitalized for the wound and underwent surgery and bed rest. The patient's chart shows that there was a smaller wound prior to the car accident, but that it got bigger after the accident. The chart also shows that there was undermining and tunneling under the surface of the wound. The patient would drive to and from florida twice a year, a 30 hour drive each way, and did not get out of the car during the trips. The patient's chart also indicated that she had been told numerous times to make sure her cushion was inflated properly and not under-inflated like she tends to have it. A therapist had pressure mapped the patient and the record shows the use of a high profile roho cushion. The nurse did not give any additional information on the cushion model, manufacture date, how it was used, etc. She was also unable to give any patient identifying information. The attorney that the nurse is working for is trying to determine if the wound was caused by the accident or if it was caused by the cushion going flat. The nurse stated she thinks the wound is most likely a result of traveling for long hours and deciding to sit during travel without performing a pressure relief. In the second conversation with the nurse, she provided the following additional information: the car accident happened on (b)(6) 2011. At the time of the accident, the patient went to the emergency room and was discharged with no injuries. On (b)(6) 2011, the patient had surgery on the wound and was placed on bed rest to allow it to heal. In (b)(6) 2011, the patient traveled to (b)(6) from (b)(6). It wasn't until the spring of 2012 that the patient claimed that the valve on the roho cushion broke at the time of the car accident
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1419507-2015-00002
MDR Report Key5022220
Report SourceHEALTH PROFESSIONAL
Date Received2015-08-11
Date of Report2015-08-11
Date of Event2011-07-30
Date Mfgr Received2015-07-14
Date Added to Maude2015-08-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. ROBIN GERGEN
Manufacturer Street100 NORTH FLORIDA AVE.
Manufacturer CityBELLEVILLE IL 622215429
Manufacturer CountryUS
Manufacturer Postal622215429
Manufacturer Phone8003562990
Manufacturer G1ROHO, INC.
Manufacturer Street100 NORTH FLORIDA AVE.
Manufacturer CityBELLEVILLE IL 622215429
Manufacturer CountryUS
Manufacturer Postal Code622215429
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameROHO HIGH PROFILE CUSHION
Product CodeKIC
Date Received2015-08-11
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROHO, INC.
Manufacturer Address100 NORTH FLORIDA AVE. BELLEVILLE IL 622215429 US 622215429


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2015-08-11

© 2024 FDA.report