NAVIOS FLOW CYTOMETER SYSTEM A52103

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2015-08-27 for NAVIOS FLOW CYTOMETER SYSTEM A52103 manufactured by Beckman Coulter Ireland.

Event Text Entries

[24622392] The customer was guided through troubleshooting procedures over the phone by a customer technical support (cts) representative. The customer ran flow check standardization beads and the fluorescent peaks were far to the right of the histogram. The cts asked the user to check if the sheath filter had air and discovered that the sheath filter was filled with 1/3 fluid and the rest air. The cts suggested to run idle mode back and forth several times to fill the sheath filter with fluid. The filter filled with fluid, and the flow check beads came within acceptable range. The instrument is operational and no service visit was required. (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[24622393] The customer reported a sheath fluid leak of approximately one liter (1 l) from the sheath filter on a navios flow cytometer system during instrument startup. No error messages were reported at the time of the event. The customer was wearing personal protective equipment (ppe) consisting of laboratory coat, goggles and gloves at the time of the event and there was no report of injury or biohazard exposure to open wounds or mucous membranes. Erroneous patient results were not generated and there was no change or effect to patient treatment in connection to the event. There was no impact to patient results and controls.
Patient Sequence No: 1, Text Type: D, B5


[39314406] The issue in this event occurred on a newly installed device that was being used by the customer for the first time; no patient samples had yet been run on the instrument. During the installation process, the sheath filter is installed onsite by a field service engineer (fse) performing the installation. The mechanism of installation is a quick disconnect, which must be fully inserted into the tubing to provide a fluid tight seal. The instrument's installation service record was reviewed and the record does not indicate any issues that occurred related to the filter installation. The likely cause of this issue is an error by the fse performing the installation consisting of a failure to fully insert the quick disconnect. There is no evidence that this issue was related to the instrument manufacturing process.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1061932-2015-01367
MDR Report Key5036246
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2015-08-27
Date of Report2015-08-03
Date of Event2015-08-03
Date Mfgr Received2016-02-22
Device Manufacturer Date2015-05-01
Date Added to Maude2015-08-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. MIRANDA HOLLAND
Manufacturer Street11800 SW 147TH AVENUE, 32-S08
Manufacturer CityMIAMI FL 331962031
Manufacturer CountryUS
Manufacturer Postal331962031
Manufacturer Phone3053802031
Manufacturer G1BECKMAN COULTER
Manufacturer Street11800 SW 147TH AVENUE
Manufacturer CityMIAMI FL 33196
Manufacturer CountryUS
Manufacturer Postal Code33196
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameNAVIOS FLOW CYTOMETER SYSTEM
Generic NameFLOW CYTOMETRIC REAGENTS AND ACCESSORIES
Product CodeOYE
Date Received2015-08-27
Model NumberNA
Catalog NumberA52103
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER IRELAND
Manufacturer AddressLISMEEHAN O'CALLAGHAN'S MILLS, COUNTY CLARE EI


Patients

Patient NumberTreatmentOutcomeDate
10 2015-08-27

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.