JUPITER SM OPERATING TABLE 4003171

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-08-27 for JUPITER SM OPERATING TABLE 4003171 manufactured by Trumpf Medizin Systeme Gmbh + Co. Hg.

Event Text Entries

[25143723] According to the investigation, the cause of the table running on its own was damage to the keypad in the remote. This issue is dealt with in the ongoing fsca ((b)(4)). The keypad used was prior to the design change. Trumpf medical has issued safety instructions to the users in the field safety notice and those should be followed until trumpf medical is able to do the final exchange. The operation table and the remote control were checked by service and repaired. Anticipated completion of the fsca is july 2016.
Patient Sequence No: 1, Text Type: N, H10


[25143724] Trumpf medical received information stating that a jupiter or trusystem 7500 operating table moved independently and without the user deliberately pressing an operating element. No patients or users were injured. This report was filed in our complaint handling system.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9681407-2015-00009
MDR Report Key5037543
Date Received2015-08-27
Date of Report2015-08-04
Date of Event2015-08-04
Date Mfgr Received2015-08-04
Date Added to Maude2015-08-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. STEFFEN ULBRICH
Manufacturer StreetCARL-ZEISS STRASSE 7-9
Manufacturer CitySAALFELD, 07318
Manufacturer CountryGM
Manufacturer Postal07318
Manufacturer Phone9367158641
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameJUPITER SM OPERATING TABLE
Generic NameBATTERY POWERED ELECTROMECHANICAL OPERATING TABLE
Product CodeFWX
Date Received2015-08-27
Model Number4003171
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerTRUMPF MEDIZIN SYSTEME GMBH + CO. HG
Manufacturer AddressCARL-ZEISS STRASSE 7-9 SAALFELD, 07318 GM 07318


Patients

Patient NumberTreatmentOutcomeDate
10 2015-08-27

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.