ADVIA CHEMISTRY TOTAL BILIRUBIN_2 REAGENT 10341115

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-08-28 for ADVIA CHEMISTRY TOTAL BILIRUBIN_2 REAGENT 10341115 manufactured by Siemens Healthcare Diagnostics Inc..

Event Text Entries

[25145059] Siemens customer care center (ccc) has notified the customer that the advia chemistry total bilirubin_2 assay has not been validated with neonatal samples and is not approved by the fda for testing with neonatal samples. The instrument is performing according to specifications. No further evaluation of the device is required.
Patient Sequence No: 1, Text Type: N, H10


[25145060] The customer is running neonatal samples for total bilirubin_2 (tbil_2) on an advia chemistry 1800 instrument. A neonatal sample run on the advia chemistry 1800 instrument gave a high result. As per the laboratory's protocol all high samples are tested on an alternate platform. The result on the alternate platform was lower. The initial result was reported to the physician(s). The result obtained on the alternate platform was not reported to the physician(s). There were no reports of adverse health consequences due to the higher result for total bilirubin on the patient sample.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2432235-2015-00380
MDR Report Key5038961
Date Received2015-08-28
Date of Report2015-08-12
Date of Event2015-08-10
Date Mfgr Received2015-08-12
Date Added to Maude2015-08-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactAARTI AZIZ
Manufacturer Street511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242683
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS (REG# 2517506)
Manufacturer Street500 GBC DRIVE P.O BOX 6101
Manufacturer CityNEWARK DE 197146101
Manufacturer CountryUS
Manufacturer Postal Code197146101
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameADVIA CHEMISTRY TOTAL BILIRUBIN_2 REAGENT
Generic NameADVIA CHEMISTRY TOTAL BILIRUBIN_2 REAGENT
Product CodeJFM
Date Received2015-08-28
Model NumberADVIA CHEMISTRY TOTAL BILIRUBIN_2 REAGENT
Catalog Number10341115
Lot Number347801
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Address511 BENEDICT AVENUE TARRYTOWN NY 10591 US 10591


Patients

Patient NumberTreatmentOutcomeDate
10 2015-08-28

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.