[24586642] Device not returned to manufacturer.
Patient Sequence No: 1, Text Type: N, H10


[24586643] On 06. 08. 2015 a usa customer reported tsh results as immeasurable using ref. (b)(4), vidas tsh 60 tests. Customer ran several other assays which came out as expected. There is no indication or report from the hospital or treating physician to biomerieux that the discrepant vidas tsh results led to any adverse event related to the patient's state of health. Recurrence of the event may potentially cause or contribute to an adverse event..
Patient Sequence No: 1, Text Type: D, B5


[35373471] Internal investigation was conducted. Review of the quality product laboratory batch records indicates vidas? Tsh batch 160226-0/1003835280 passed all performance testing; no anomalies were cited. This lot performed in accordance with specifications. The customer submittals (3) were tested via the following methods: vidas? Tsh, vidas? Tsh3, vidas? Ft4, cobas tsh/roche. The results of all methods consistently indicate hyperthyro? Dism; the inconsistent results obtained by the customer are not reproduced/confirmed. The vidas? Tsh ref. (b)(4) batch 1003835280/160223-0 is performing within specifications.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002769706-2015-00077
MDR Report Key5041657
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2015-08-31
Date of Report2015-08-06
Date of Event2015-08-06
Device Manufacturer Date2015-03-11
Date Added to Maude2015-08-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. RYAN LEMELLE
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147318582
Manufacturer G1BIOMERIEUX SA
Manufacturer StreetCHEMIN DE L ORME
Manufacturer CityMARCY L ETOILE, RHONE 69280
Manufacturer CountryFR
Manufacturer Postal Code69280
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVIDAS? TSH ASSAY
Generic NameVIDAS? TSH ASSAY
Product CodeLYR
Date Received2015-08-31
Catalog Number30400-01
Lot Number1003835280
Device Expiration Date2016-02-26
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer AddressCHEMIN DE L ORME MARCY L ETOILE, RHONE 69280 FR 69280


Patients

Patient NumberTreatmentOutcomeDate
10 2015-08-31