MAVIG LEAD SHIELD 0T25U01 *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2004-01-05 for MAVIG LEAD SHIELD 0T25U01 * manufactured by Ti-ba Enterprises.

Event Text Entries

[18560526] Patient was on the cath lab table (room 1) and the procedure was approx 2/3 complete when the lead shield suddenly broke from its support arm and crashed to the floor. No one was hurt, but the shield did come close to falling on the pt and the physician performing the procedure. The lead shield was in a stationary position at the time of the event and according to the staff had been moved for positioning two times during the procedure up to the point of breaking off. The ge representative, happened to be in the cath lab area doing maintenance on cath room #2. The incident was reported to them as well. Pt does not know of any history associated with other incidents with this product. It appears that the extension arm of the shield broke at a "weld point" on the first elbow of the swing arm. Device usage problem: device failed (e. G. Broke, couldn't get it to work or stopped working). Other devices in use on patient: no, the patient was on the cath lab table while the cardiologist and team were doing a cath procedure on the patient. The lead shield broke from its support arm and fell to the floor, missing the patient and staff member nearby.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number504190
MDR Report Key504190
Date Received2004-01-05
Date of Report2003-10-01
Date of Event2003-10-01
Date Added to Maude2004-01-06
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameMAVIG LEAD SHIELD
Generic NameLEAD SHIELD USED ON G.E. BI-PLANE CATH UNIT
Product CodeKPY
Date Received2004-01-05
Model Number0T25U01
Catalog Number*
Lot Number*
ID Number*
Device AvailabilityY
Device Age28 MO
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key492982
ManufacturerTI-BA ENTERPRISES
Manufacturer Address25 HYTEC CIRCLE ROCHESTER NY 14606 US