TC MAYO-HEGAR NDL HOLDERHVYSERR185MM BM066R

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2015-08-13 for TC MAYO-HEGAR NDL HOLDERHVYSERR185MM BM066R manufactured by Aesculap Ag & Co. Kg.

Event Text Entries

[25060588] Manufacturing site evaluation: evaluation on-going.
Patient Sequence No: 1, Text Type: N, H10


[25060589] Facility reports the device broke during surgery and a piece was lost inside the patient.
Patient Sequence No: 1, Text Type: D, B5


[35883601] Manufacturing site evaluation: product received with one of the hard metal inlays is broken 2 mm from the tip. Investigation was completed visually, using the digital microscope. This failure is user/maintenance related. The tip of the carbide metal may have broken off due to a needle being picked up and it slipped off the needle holder or by aggressive maintenance methods, the solder area may have dissolved, resulting in the carbide metal being brittle. Corrective / preventive action(s): not applicable.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2916714-2015-00751
MDR Report Key5046424
Report SourceCOMPANY REPRESENTATIVE
Date Received2015-08-13
Date of Report2016-01-13
Date Mfgr Received2015-07-24
Date Added to Maude2015-09-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNICOLE BROYLES
Manufacturer Street615 LAMBERT POINTE DRIVE
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3145515988
Manufacturer G1AESCULAP AG&CO KG
Manufacturer StreetPO BOX 40
Manufacturer CityTUTTLINGEN 78501
Manufacturer CountryGM
Manufacturer Postal Code78501
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTC MAYO-HEGAR NDL HOLDERHVYSERR185MM
Generic NameNEEDLE HOLDER
Product CodeHXK
Date Received2015-08-13
Model NumberBM066R
Catalog NumberBM066R
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerAESCULAP AG & CO. KG
Manufacturer AddressTUTTLINGEN DE 78532 DE 78532


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-08-13

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.