MST MALYUGIN RING MAL-0002-1 MAL-0002

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-09-01 for MST MALYUGIN RING MAL-0002-1 MAL-0002 manufactured by Microsurgical Technology Inc.

Event Text Entries

[25387272] The ring was returned with a fractured glue joint and showing signs of being squeezed or pinched during use. The physician indicated that he was using a method of removal that is contrary to the instructions for use. The doctor imployed the use of the osher malyugin manipulator rather than the injector to remove the ring from the eye.
Patient Sequence No: 1, Text Type: N, H10


[25387273] The surgeon reported that when he attempted to remove the malyugin ring from the patient's eye with the osher/malyugin manipulator the ring broke at the glue joint. There was no impact to the patient and the ring was removed from the eye.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3019924-2015-00028
MDR Report Key5048468
Date Received2015-09-01
Date of Report2015-09-01
Date of Event2015-08-03
Date Mfgr Received2015-08-05
Date Added to Maude2015-09-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. ROBERT MAY
Manufacturer Street8415 154TH AVE NE
Manufacturer CityREDMOND WA 98052
Manufacturer CountryUS
Manufacturer Postal98052
Manufacturer Phone4255560544
Manufacturer G1MICROSURGICAL TECHNOLOGY INC
Manufacturer Street8415 154TH AVE NE
Manufacturer CityREDMOND WA 98052
Manufacturer CountryUS
Manufacturer Postal Code98052
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameMST MALYUGIN RING
Generic NameIRIS RETRACTOR CLIP
Product CodeHOC
Date Received2015-09-01
Returned To Mfg2015-08-07
Model NumberMAL-0002-1
Catalog NumberMAL-0002
OperatorPHYSICIAN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMICROSURGICAL TECHNOLOGY INC
Manufacturer Address8415 154TH AVE NE REDMOND WA 98052 US 98052


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-09-01

© 2026 FDA.report