[24794579] The complete device was not returned to pelton & crane as the pelton & crane distributor had already repaired the dental light by replacing the counterbalance spring assembly. However, during follow-up communication with the doctors office, the replaced counterbalance spring assembly was still in the dr. 's possession therefore pelton & crane had the part returned for evaluation. Pelton & crane received the counterbalance spring assembly august 19, 2015. Upon evaluation of the spring assembly, it was identified that the two pivot studs located on both sides of the friction block were broken off due to suspected excessive loading therefore the spring assembly could not counterbalance the light head assembly. The pelton & crane dental track light was manufactured twenty-six (26) years ago in august 1989 and is past the life of the device. Furthermore, dr. (b)(6) purchased the dental track light used in 2006 from another dentist.
Patient Sequence No: 1, Text Type: N, H10


[24794580] A dental hygienist was performing routine dental treatment with a patient when a pelton & crane lftii dental light counter balance arm assembly failed causing the light arm to drop down towards the ground. The light arm dropped approximately one to two feet then struck the dental assistant on the right side of her head. The dental assistant went to the hospital due to feeling pain and was diagnosed with a fracture of the occipital bone and a contusion. The dental hygienist stayed in the hospital for two nights and was then released.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1017522-2015-00019
MDR Report Key5048942
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2015-09-02
Date of Report2015-08-05
Date of Event2015-07-08
Date Mfgr Received2015-08-05
Device Manufacturer Date1989-08-31
Date Added to Maude2015-09-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. FRANK RAY
Manufacturer Street11727 FRUEHAUF DRIVE
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal28273
Manufacturer Phone7045877227
Manufacturer G1PELTON & CRANE
Manufacturer Street11727 FRUEHAUF DRIVE
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal Code28273
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLIGHT FANTASTIC II TRACK MOUNT LIGHT
Generic NameDENTAL LIGHT
Product CodeEAZ
Date Received2015-09-02
Model NumberHBT
Catalog NumberLFTII
OperatorDENTAL HYGIENIST
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerPELTON & CRANE
Manufacturer Address11727 FRUEHAUF DRIVE CHARLOTTE NC 28273 US 28273


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Required No Informationntervention 2015-09-02