VIDAS ANALYZER 99735

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-09-02 for VIDAS ANALYZER 99735 manufactured by Biom?rieux Italia.

Event Text Entries

[25503650] Device not returned to manufacturer.
Patient Sequence No: 1, Text Type: N, H10


[25503651] A customer in the united states reported qcv failure on slot in association with the vidas? Analyzer ((b)(4)). The customer repeated the qcv test; the same failure occurred. The last successful qcv was performed (b)(6) 2015. The local field service engineer (fse) performed a site evaluation on (b)(6) 2015 and reported a seal required cleaning. Following fse actions, a subsequent qcv run was performed and passing results indicated the system is performing to specifications. There was no reported death or serious injury associated with the identified event. Internal biomerieux investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


[36737890] Biom? Rieux investigation was performed at the customer site by the local field service engineer (fse). The qcv failure was reproduced in section by the fse. Investigation identified debris in section. The fse removed the debris and performed qualification tests. Following repair, the qcv test passes; the section is performing in accordance with specifications. The vidas? System was returned to operation.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002769706-2015-00079
MDR Report Key5051495
Date Received2015-09-02
Date of Report2015-08-21
Date Mfgr Received2015-08-21
Date Added to Maude2015-09-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. RYAN LEMELLE
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147318582
Manufacturer G1BIOMERIEUX, INC
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityST. LOUIS MO 63042
Manufacturer CountryUS
Manufacturer Postal Code63042
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVIDAS ANALYZER
Generic NameVIDAS ANALYZER
Product CodeDEW
Date Received2015-09-02
Model Number99735
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOM?RIEUX ITALIA
Manufacturer AddressVIA DI CAMPIGLIANO 58 PONTE-A EMA FIRENZE, ITALY 50015 IT 50015


Patients

Patient NumberTreatmentOutcomeDate
10 2015-09-02

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.