[26484866] A siemens customer service engineer (cse) specialist was dispatched to the customer site. After evaluating the instrument, the cse found a stray cuvette had fallen underneath the luminometer waste probe and was removed. The cse performed an empty and fill and the customer performed a daily cleaning. The cause of a delay in testing was related to the cuvette jam. The instrument is performing according to specifications. No further evaluation of the device is required.
Patient Sequence No: 1, Text Type: N, H10


[26484867] The operator of an advia centaur xp instrument reported a delay in testing of an infectious disease methods. There are no known reports of patient intervention or adverse health consequences due to the delay in testing of an infectious disease methods.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2432235-2015-00394
MDR Report Key5056767
Date Received2015-09-04
Date of Report2015-08-09
Date of Event2015-08-09
Date Mfgr Received2015-08-09
Device Manufacturer Date2012-04-09
Date Added to Maude2015-09-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCASSANDRA KOCSIS
Manufacturer Street511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242687
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS
Manufacturer StreetMANUFACTURING LIMITED REGISTRATION NUMBER: 8020888
Manufacturer CityCHAPEL LANE, SWORDS, DUBLIN
Manufacturer CountryEI
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameADVIA CENTAUR XP
Generic NameIMMUNOASSAY ANALYZER
Product CodeLOB
Date Received2015-09-04
Model NumberADVIA CENTAUR XP
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No0
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Address511 BENEDICT AVE TARRYTOWN NY 10591 US 10591


Patients

Patient NumberTreatmentOutcomeDate
10 2015-09-04

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.