ELEVATOR 3714383 GORNEY SUCTION 3MM TIP

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-09-09 for ELEVATOR 3714383 GORNEY SUCTION 3MM TIP manufactured by Integra York, Pa Inc..

Event Text Entries

[25914483] To date the device involved in the reported incident has not been received for evaluation. An investigation has been initiated based on the reported information.
Patient Sequence No: 1, Text Type: N, H10


[25914484] During endoscopic sinus surgery the part broke in patients nose. There was no harm to the patient and the parts were retrieved.
Patient Sequence No: 1, Text Type: D, B5


[32666271] On 10/7/2015 integra investigation completed. Device history evaluation. Failure analysis: there was a gorney suction elevator returned; not showing any unusual markings. The suction elevator returned in used condition showing wear. Evaluation of the suction elevator shows that the shaft/tube is broken off at before the bulb. The complaint report is confirmed. Device history evaluation : dhr complete with all available history. Nonconforming product report / nonconforming material report history: none. Variance authorization / deviation history: none. Engineering change order/manufacturing change order history: none. Corrective action preventive action history: none. Health hazard evaluation history: none. Conclusion: the root cause has not been identified as a workmanship or material deficiency.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2523190-2015-00073
MDR Report Key5063710
Date Received2015-09-09
Date of Report2015-08-19
Date of Event2015-08-19
Date Mfgr Received2015-10-07
Date Added to Maude2015-09-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactUSER SANDRA LEE
Manufacturer Street311 ENTERPRISE DRIVE
Manufacturer CityPLAINSBORO NJ 08536
Manufacturer CountryUS
Manufacturer Postal08536
Manufacturer Phone6099362393
Manufacturer G1INTEGRA YORK, PA INC.
Manufacturer Street589 DAVIES DRIVE
Manufacturer CityYORK PA 17402
Manufacturer CountryUS
Manufacturer Postal Code17402
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameELEVATOR 3714383 GORNEY SUCTION 3MM TIP
Generic NamePFM11
Product CodeKAD
Date Received2015-09-09
Returned To Mfg2015-08-20
Catalog Number3714383
Lot NumberUNKNOWN
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerINTEGRA YORK, PA INC.
Manufacturer Address589 DAVIES DRIVE 589 DAVIES DRIVE YORK PA 17402 US 17402


Patients

Patient NumberTreatmentOutcomeDate
10 2015-09-09