STETH, BOWLES, BK SES05ABK

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2015-09-11 for STETH, BOWLES, BK SES05ABK manufactured by Honsun Nantong Company Ltd..

Event Text Entries

[25623622] (b)(4). The device history record could not be reviewed as a lot number was not provided. Multiple attempts to obtain samples from the customer were made and no sample has been received. Review of current production records did not indicate any exception that could lead to the reported incident. A review of the ear tip design demonstrated that the ear bud was secured firmly to the stethoscope and would not fall off accidentally based on the following: the ear tip inner diameter dimension is 4+/-0. 05mm, the binaural od dimension is 4. 9+0. 05mm, providing a tight interface between the ear tip and binaural. Test conducted on the assembly indicated a 3. 0~4. 0kgf force was needed to pull the ear tip off from binaural. It is unlikely that the ear tip would drop off without intentional removal effort. A review of the assembly process was conducted and confirmed to be in control. No records of improper assembly were identify during the investigation. Without the return of the complaint sample for further investigation, the root cause of the missing ear bud could not be determined in this investigation. No additional action will be taken at this time. We will continue to monitor for this type of incident. F
Patient Sequence No: 1, Text Type: N, H10


[25623623] Customer reported that the rubber ear bud had fallen off the stethoscope and the nurse did not realize it was missing. He (the nurse) placed the stethoscope in his ears and the side without the ear bud went all the way in his ear and pierced his ear drum. Multiple attempts to obtain further patient information have been made but hospital states they are unable to provide.
Patient Sequence No: 1, Text Type: D, B5


[34031288] (b)(4). The actual sample was returned for investigation on sep 22, 2015. The device history record review did not indicate any exception that could lead to the reported incident. Inspection conducted indicated one missing ear tip but no abnormality was observed. Tension force test on the side with the ear tip indicated an acceptable force of 3. 15kgf. From the investigation, the root cause of the missing ear bud could not be determined. No additional action will be taken at this time. We will continue to monitor for this type of event.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1423537-2015-00062
MDR Report Key5070815
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2015-09-11
Date of Report2015-09-11
Date of Event2015-08-13
Date Facility Aware2015-08-14
Date Mfgr Received2015-08-13
Date Added to Maude2015-09-11
Event Key0
Report Source CodeDistributor report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMICHELE DONATICH
Manufacturer Street1500 WAUKEGAN ROAD
Manufacturer CityWAUKEGAN IL 60085
Manufacturer CountryUS
Manufacturer Postal60085
Manufacturer Phone8478876412
Manufacturer G1HONSUN NANTONG COMPANY LTD.
Manufacturer StreetNO.8 TONGXING ROAD ECONOMIC&TECHNICAL DEVELOPMENT
Manufacturer CityNANTONG CITY
Manufacturer CountryCH
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTETH, BOWLES, BK
Generic NameSTETH, BOWLES, BK
Product CodeLDE
Date Received2015-09-11
Model NumberSES05ABK
Catalog NumberSES05ABK
Lot NumberNO LOT GIVEN
OperatorNURSE
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerHONSUN NANTONG COMPANY LTD.
Manufacturer AddressNO.8 TONGXING ROAD ECONOMIC&TECHNICAL DEVELOPMENT NANTONG CITY CH


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-09-11